Beckman Coulter, Inc. recalls
46 notices naming Beckman Coulter, Inc. as the recalling firm, since December 18, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Beckman Coulter, Inc.: ROTOR ASSY-SW40 and 1 more
Hazard (FDA): Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being…
FDAMedical devices
Beckman Coulter, Inc.: DxU 840m Iris and 1 more
Hazard (FDA): The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual…
FDAMedical devices
Beckman Coulter Inc.: DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE and 1 more
Hazard (FDA): Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system…
FDAMedical devices
Beckman Coulter Inc.: Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower… and 1 more
Hazard (FDA): a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a…
FDAMedical devices
Beckman Coulter Inc.: REF # D23352 and 1 more
Hazard (FDA): Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
FDAMedical devices
Beckman Coulter, Inc.: Access Total T4 Calibrator
Hazard (FDA): Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a…
FDAMedical devices
Beckman Coulter, Inc.: MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended…
Hazard (FDA): Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
FDAMedical devices
Beckman Coulter, Inc.: MicroScan Neg MIC 3J REF C54814
Hazard (FDA): Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
FDAMedical devices
Contamination
Beckman Coulter, Inc.: MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are…
Hazard (FDA): Due to the likely presence of contamination in well(s).
FDAMedical devices
Beckman Coulter, Inc.: CellMek SPS Sample Preparation System
Hazard (FDA): Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in…
FDAMedical devices
Beckman Coulter, Inc.: COULTER DxH Diluent
Hazard (FDA): Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
FDAMedical devices
Beckman Coulter, Inc.: Beckman Coulter UniCel Dxl 600 and 1 more
Hazard (FDA): Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs…
FDAMedical devices
Beckman Coulter, Inc.: Access 2 Reaction Vessels
Hazard (FDA): It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could…
FDAMedical devices
Crash
Beckman Coulter Inc.: DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow…
Hazard (FDA): Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
FDAMedical devices
Beckman Coulter, Inc.: Beckman Coulter UniCel DxI 600 and 1 more
Hazard (FDA): An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading…
FDAMedical devices
Beckman Coulter Inc.: SYNCHRON Systems Direct LDL Cholesterol
Hazard (FDA): Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally…
FDAMedical devices
Beckman Coulter, Inc.: CellMek SPS Sample Preparation System
Hazard (FDA): Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters…
FDAMedical devices
Beckman Coulter, Inc.: DxC 500i Clinical Analyzer Modules and 1 more
Hazard (FDA): Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering…
FDAMedical devices
Beckman Coulter Inc.: SYNCHRON Systems Phosphorus
Hazard (FDA): Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
FDAMedical devices
Beckman Coulter Inc.: DxC 500 AU Module w/ISE
Hazard (FDA): On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress…
FDAMedical devices
Beckman Coulter Inc.: UniCel DxH 900 COULTER Cellular Analysis System and 3 more
Hazard (FDA): Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
FDAMedical devices
Beckman Coulter, Inc.: Access Thyroglobulin assay
Hazard (FDA): Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or…
FDAMedical devices
Beckman Coulter Inc.: UIBC
Hazard (FDA): Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL…
FDAMedical devices
Beckman Coulter, Inc.: UniCel DxI 800 Access Immunoassay Analyzer and 5 more
Hazard (FDA): Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current…
FDAMedical devices
- FDABeckman Coulter Inc.: Estrone RIA
- FDABeckman Coulter, Inc.: Access Vitamin B12 Reagent
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer C11137
- FDABeckman Coulter Inc.: REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter, Inc.: Access Testosterone assay
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzer
- FDABeckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzer
- FDABeckman Coulter, Inc.: Beckman Coulter DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzer
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter, Inc.: Access 2 Immunoassay Analyzer
- FDABeckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter, Inc.: Access Erythropoietin
- FDABeckman Coulter, Inc.: Beckman Coulter DxI 9000 Access Immunoassay Analyzer
- FDABeckman Coulter Inc.: Access PCT Reagent Pack
- FDABeckman Coulter, Inc.: Access Intact PTH assay
- FDABeckman Coulter, Inc.: Access hsTnI Reagent
- FDABeckman Coulter Inc.: Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer