Why it was recalled

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Access 2 Immunoassay Analyzer, Catalog Number 81600N

    Lot or code
    UDI-DI: 15099590265380; Serial Numbers: 574626 574627 574839 574840 574841 574842 574852 574853 574854 574378 573724 574432 574538 574907 574497 574303 574184 574218 574219 574305 574332 574333 574345 574346 574414 574451 574581 574582 574583 574584 574726 574727 574728 574729 574730 574731 574732 574733 574920 574922 574923 574955 574957 574958 574959 574984 574985 574986 574987 574811 574160 574161 574167 574168 574169 574171 574177 574178 574179 574180 574182 574230 574297 574300 574465 574466 574467 574468 574615 574735 574807 574858 574887 574893 574896 574912 574913 574925 574926 574950 574952 574979 574982 573109 573073 573072 574585 574586 574587 574588 574589 574590 574172 574173 574174 574175 574181 574183 574224 574227 574238 574239 574244 574245 574253 574254 574256 574258 574259 574267 574268 574269 574270 574271 574284 574295 574296 574313 574335 574336 574338 574392 574393 574394 574395 574396 574397 574398 574399 574400 574402 574403 574404 574405 574406 574407 574520 574536 574576 574577 574578 574579 574580 574591 574616 574618 574619 574620 574621 574675 574676 574677 574678 574681 574682 574683 574684 574685 574741 574829 574830 574831 574832 574865 574866 574867 574869 574879 574880 574881 571672 510520 512247 573221 575002 574225 574401 574194 574199 574371 574773 574774 574093 574094 574129 574202 574204 574211 574252 574257 574352 574201 574251 574261 574369 574476 574478 574495 574628 509167 570853 574330 574331 574469 574470 574471 574209 574213…
    FDA recall number
    Z-1386-2025
    Quantity
    873 units

    The text above is the beginning of the record; 4,655 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Beckman Coulter, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Beckman Coulter, Inc., Month: March 2025, Week 2025-W13. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96248 on March 26, 2025.