What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.
- Lot or code
- UDI/DI 15099590693183, All Lots
- FDA recall number
- Z-0727-2025
- Quantity
- 2,109,804 units
Distribution (FDA): Worldwide
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beckman Coulter, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.