Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96876: Beckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer C11137

FDA recall 96876 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

239 more codes are in the record below.

  • DxI 9000 Access Immunoassay Analyzer C11137

    Lot or code
    UDI-DI 15099590732103 Serial Numbers 300116 300124 300126 300128 300132 300136 300138 300144 300145 300148 300150 300153 300154 300155 300156 300157 300158 300161 300166 300168 300171 300174 300175 300178 300179 300182 300183 300186 300190 300191 300192 300193 300194 300198 300200 300201 300204 300205 300206 300207 300208 300209 300210 300211 300212 300213 300214 300215 300217 300218 300219 300221 300225 300226 300228 300231 300232 300234 300235 300236 300240 300242 300246 300249 300250 300252 300255 300266 300267 300270 300272 300273 300274 300275 300279 300280 300282 300283 300284 300285 300286 300288 300293 300299 300300 300306 300307 300308 300309 300312 300314 300316 300317 300318 300319 300321 300322 300323 300326 300327 300328 300329 300331 300333 300334 300336 300338 300339 300341 300346 300350 300352 300354 300355 300356 300358 300359 300360 300361 300362 300370 300372 300373 300375 300376 300378 300388 300389 300392 300393 300395 300396 300397 300398 300403 300407 300408 300409 300410 300411 300413 300414 300415 300420 300422 300423 300424 300425 300426 300427 300428 300430 300434 300438 300446 300454 300455 300457 300458 300461 300464 300467 300468 300470 300474 300475 300476 300479 300481 300483 300489 300490 300491 300493 300495 300496 300497 300500 300501 300506 300507 300508 300509 300510 300511 300512 300514 300515 300516 300517 300518 300519 300521 300522 300523 300525 300526 300527 300528 300530 300531 300532 300533 300534 300535 300536 300537 300538 300539…
    FDA recall number
    Z-1996-2025
    Quantity
    67 US; 178 OUS

    The text above is the beginning of the record; 259 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Libya, New Zealand, Oman, Poland, Portugal, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.

Source

Issued by
U.S. Food and Drug Administration
Firm
Beckman Coulter, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Beckman Coulter, Inc., Month: July 2025, Week 2025-W27. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96876 on July 2, 2025.

Other notices of Beckman Coulter, Inc.