Why it was recalled

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

391 more codes are in the record below.

  • Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

    Lot or code
    A71460, UDI/DI 15099590369224, Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583; A30260, UDI/DI 15099590341602, Serial Numbers: 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576.
    FDA recall number
    Z-0428-2026
    Quantity
    80 units
  • Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer

    Lot or code
    973100, UDI/DI 15099590265373 (Chaska), Serial numbers: 610458, 610455, 610679, 610579, 610432, 610430, 610434, 610664, 610682, 610765, 610500, 610541, 610544, 610545, 610562, 610547, 610585, 610595, 610475, 610453, 610781, 610580, 610444, 610704, 610586, 610684, 610589, 610588, 610493, 610659, 610570, 610498, 610748, 610496, 610560, 610573, 610574, 610577, 610601, 610651, 610680, 610642, 610641, 610683, 610747, 610468, 610772, 610441, 610476, 610489, 610488, 610492, 610532, 610534, 610533, 610535, 610640, 610457, 610464, 610790, 610559, 610687, 610571, 610552, 610555, 610582, 610531, 610568, 610594, 610773, 610490, 610668, 610784, 610812, 610556, 610550, 610600, 610565, 610558, 610557, 610551, 610527, 610685, 610730, 610686, 610494, 610753, 610590, 610593, 610546, 610724, 610538, 610540, 610439, 610491, 610581, 610649, 610466, 610789, 610723, 610575, 610443, 610776, 610734, 610518, 610519, 610517, 610520, 610563, 610569, 610787, 610459, 610775, 610572, 610561, 610597, 610445, 610440, 610567, 610596, 610497, 610774, 610454, 610426, 610431, 610442, 610433, 610435, 610539, 610578, 610509, 610418, 610483, 610484, 610515, 610425, 610503, 610510, 610452, 610526, 610602, 610481, 610487, 610482, 610507, 610505, 610506, 610514, 610499, 610504, 610511, 610423, 610524, 610525, 610512, 610564, 610424, 610513, 610480, 610516, 610521, 610419, 610415, 610416, 610456, 610461, 610460, 610472, 610470, 610477, 610471, 610463, 610462, 610467, 610469, 610465, 610587, 610639, 610614, 610616,…
    FDA recall number
    Z-0429-2026
    Quantity
    657 units

    The text above is the beginning of the record; 3,032 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide

Source

Issued by
U.S. Food and Drug Administration
Firm
Beckman Coulter, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Beckman Coulter, Inc., Month: November 2025, Week 2025-W46. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97730 on November 12, 2025.

Other notices of Beckman Coulter, Inc.