What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860
- Lot or code
- UDI-DI: 15099590227173; Lot Number: 439163
- FDA recall number
- Z-2322-2025
- Quantity
- 453 units
Distribution (FDA): Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy, Malaysia, Mexico, Philippines, Spain.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beckman Coulter, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.