What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
- Lot or code
- UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C
- FDA recall number
- Z-2152-2025
- Quantity
- 496 units
Distribution (FDA): US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beckman Coulter Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.