Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96273: Beckman Coulter, Inc.: DxI 9000 Access Immunoassay Analyzer

FDA recall 96273 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

239 more codes are in the record below.

  • DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

    Lot or code
    UDI-DI: 15099590732103; Serial Numbers: 300116 300117 300123 300124 300126 300128 300129 300130 300131 300132 300133 300134 300135 300136 300137 300138 300139 300141 300142 300143 300144 300145 300147 300148 300149 300150 300151 300152 300153 300154 300155 300156 300157 300158 300161 300162 300163 300164 300166 300167 300168 300169 300170 300171 300172 300173 300174 300176 300177 300178 300179 300181 300182 300183 300186 300188 300189 300190 300191 300192 300193 300194 300195 300197 300198 300199 300200 300201 300202 300203 300204 300205 300206 300207 300208 300209 300210 300211 300212 300213 300214 300215 300216 300217 300218 300219 300220 300221 300222 300223 300224 300225 300226 300227 300228 300229 300230 300231 300232 300233 300234 300235 300236 300237 300238 300239 300240 300241 300242 300243 300244 300245 300246 300247 300248 300250 300251 300252 300253 300254 300255 300256 300257 300258 300259 300260 300261 300263 300264 300266 300267 300268 300269 300270 300271 300272 300273 300274 300275 300276 300277 300279 300280 300281 300282 300283 300284 300285 300286 300287 300288 300289 300290 300291 300293 300294 300296 300297 300298 300299 300300 300301 300302 300303 300304 300305 300306 300307 300308 300309 300312 300313 300314 300315 300316 300317 300318 300319 300320 300321 300322 300324 300325 300326 300328 300329 300330 300331 300332 300333 300334 300336 300337 300338 300339 300340 300341 300343 300346 300349 300350 300351 300352 300353 300354 300355 300356 300358…
    FDA recall number
    Z-1326-2025
    Quantity
    245 units

    The text above is the beginning of the record; 259 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): US Nationwide. Global Distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Beckman Coulter, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Beckman Coulter, Inc., Month: March 2025, Week 2025-W12. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96273 on March 19, 2025.