What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
20 more codes are in the record below.
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
- Lot or code
- UDI-DI: 15099590732103; Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352
- FDA recall number
- Z-1388-2025
- Quantity
- 206 units
Distribution (FDA): US Nationwide. Global Distribution.