What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
134 more codes are in the record below.
UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
- Lot or code
- UDI: 15099590701574/ Serial Numbers: BB33238, BB33239, BB42317, BC05150, BC08198, BC10226, BC16320, BC21392, BC23453, BC25457, BC25465, BC30527, BC32542, BC42721, BC42726, BC49882, BD13219, BD41570, BD46639, BD46643, BD46646, BE01022, BE02044, BE02052, BE04080, BE21222, BE24254, BE24257, BE29334, BE31336, BE33366, BE42455, BE42459, BE46517, BE49539, BE49541, BF10071, BF12126, BF12129, BF18194, BF33377, BF40462, BG04041, BG04042, BG04053, BG05080, BG11182, BG11184, BG11185, BG14195, BG14200, BG14204, BG14208, BG16220, BG16231, BG17238, BG17250, BG17260, BG19283, BG20292, BG20294, BG25395, BG25412, BG25413, BG26435, BG40563, BG45631, BG46635, BH04028, BH04029, BH05051, BH19220, BH24260, BH31328, BD46642, BG22347, BG20307, BD46640, BC32545
- FDA recall number
- Z-2572-2025
UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
- Lot or code
- UDI: 15099590729110/ Serial Numbers: BD17734, BE12048, BE28105, BF14045, BF23096, BF46236, BG26135, BG50247, BH12019, BH43168
- FDA recall number
- Z-2573-2025
UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
- Lot or code
- UDIs: 15099590672423, 15099590381646, 15099590687922/ Serial Numbers: AT38527, AU02008, AU10078, AV28630, AW13187, AW29123, BC18100, RBB03201, AW43276, AY06074, AY06092, AY41556, AY43605, AZ02030, AZ03041, AZ18245, AZ22329, AZ39513, AZ46681, BA03048, BA03050, BA03055, BA03060, BA07061, BA07062, BA13227, BA15245, BA27414, BB16181, BB29286, BC08036, BC18099, AZ25376, BB25262, BB25261, BA25393
- FDA recall number
- Z-2574-2025
UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
- Lot or code
- UDI: 15099590192242/ Serial Numbers: AV52143, AZ02628, AZ06676, AZ24836, BA07048, BA08667, BA17727, BA31855, BB10648, BB49881, BC24713
- FDA recall number
- Z-2575-2025
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beckman Coulter Inc. (other notices of this firm)
- Title