Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97464: Beckman Coulter Inc.: UniCel DxH 900 COULTER Cellular Analysis System and 3 more

FDA recall 97464 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

134 more codes are in the record below.

  • UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478

    Lot or code
    UDI: 15099590701574/ Serial Numbers: BB33238, BB33239, BB42317, BC05150, BC08198, BC10226, BC16320, BC21392, BC23453, BC25457, BC25465, BC30527, BC32542, BC42721, BC42726, BC49882, BD13219, BD41570, BD46639, BD46643, BD46646, BE01022, BE02044, BE02052, BE04080, BE21222, BE24254, BE24257, BE29334, BE31336, BE33366, BE42455, BE42459, BE46517, BE49539, BE49541, BF10071, BF12126, BF12129, BF18194, BF33377, BF40462, BG04041, BG04042, BG04053, BG05080, BG11182, BG11184, BG11185, BG14195, BG14200, BG14204, BG14208, BG16220, BG16231, BG17238, BG17250, BG17260, BG19283, BG20292, BG20294, BG25395, BG25412, BG25413, BG26435, BG40563, BG45631, BG46635, BH04028, BH04029, BH05051, BH19220, BH24260, BH31328, BD46642, BG22347, BG20307, BD46640, BC32545
    FDA recall number
    Z-2572-2025
  • UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520

    Lot or code
    UDI: 15099590729110/ Serial Numbers: BD17734, BE12048, BE28105, BF14045, BF23096, BF46236, BG26135, BG50247, BH12019, BH43168
    FDA recall number
    Z-2573-2025
  • UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

    Lot or code
    UDIs: 15099590672423, 15099590381646, 15099590687922/ Serial Numbers: AT38527, AU02008, AU10078, AV28630, AW13187, AW29123, BC18100, RBB03201, AW43276, AY06074, AY06092, AY41556, AY43605, AZ02030, AZ03041, AZ18245, AZ22329, AZ39513, AZ46681, BA03048, BA03050, BA03055, BA03060, BA07061, BA07062, BA13227, BA15245, BA27414, BB16181, BB29286, BC08036, BC18099, AZ25376, BB25262, BB25261, BA25393
    FDA recall number
    Z-2574-2025
  • UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858

    Lot or code
    UDI: 15099590192242/ Serial Numbers: AV52143, AZ02628, AZ06676, AZ24836, BA07048, BA08667, BA17727, BA31855, BB10648, BB49881, BC24713
    FDA recall number
    Z-2575-2025

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.

Source

Issued by
U.S. Food and Drug Administration
Firm
Beckman Coulter Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Beckman Coulter Inc., Month: September 2025, Week 2025-W39. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97464 on September 24, 2025.