Philips North America recalls
11 notices naming Philips North America as the recalling firm, since June 18, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Philips North America: Model Number: 712010.
Hazard (FDA): Potential for the device to fail immediately after an exposure, causing the system to remain stationary.
FDAMedical devices
Philips North America: Philips Smart-hopping 2.0 AP 1.4 GHz
Hazard (FDA): Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC…
FDAMedical devices
Philips North America: Ingenia Blue Seal SE and Blue Seal XE and 11 more
Hazard (FDA): When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction…
FDAMedical devices
Philips North America: Philips Patient Information Center iX REF:866389 and 1 more
Hazard (FDA): Due to software malfunction, patient information system may freeze and become unresponsive.
FDAMedical devices
Philips North America: Philips Achieva 1.5T with MR Elastography and 18 more
Hazard (FDA): The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE…
FDAMedical devices
Philips North America: Evolution Upgrade 1.5T. Product Codes and 12 more
Hazard (FDA): The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
FDAMedical devices
Philips North America: Patient Information Center iX
Hazard (FDA): A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
FDAMedical devices
Philips North America: Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL…
Hazard (FDA): Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance…
FDAMedical devices
Philips North America: Cardiac Workstation 5000 and 1 more
Hazard (FDA): Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
FDAMedical devices
Fire
Philips North America: Achieva 3.0T and 18 more
Hazard (FDA): The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
FDAMedical devices
Fire
Philips North America: Achieva XR Product Numbers: and 8 more
Hazard (FDA): identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the…
FDAMedical devices