What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
- Lot or code
- Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016
- FDA recall number
- Z-0597-2026
- Quantity
- 110
Distribution (FDA): U.S.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips North America (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.