What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
167 more codes are in the record below.
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
- Lot or code
- Product/Model Number: 867216; UDI-DI: N/A; System Serial Numbers: DK61624330, DK61623830, DK61724824, DK61623802, DK61623826, DK61624101, DK61624082, DK61623910, DK61624187, DK51313245, DK61724472, DK61724620, DK61724590, DK61724461, DK61724725, DK61724679, DK61724684, DK51313645, DK51313455, DK61724814, DK61724747, DK61624235, DK61422548, DK52316132, DK61724553, DK61724957, DK61724756, DK61523178, DK61724502, DK61724860, DK61523117, DK61724577, DK61724376, DK61724566, DK61422553, DK61322150, DK61523070, DK61523270, DK52617072, DK61523151, DK61422454, DK53817745, DK52617073, DK61422674, DK61422973, DK61523211, DK61322124, DK61523272, DK61523248, DK51313320, DK61422463, DK61523260, DK61322137, DK61322337, DK61322253, DK61322375, DK61322380, DK61422762, DK61422868, DK61322122, DK61422851, DK61322283, DK61523658, DK61422848, DK61422884, DK61422897, DK61523409, DK61422808, DK61422736, DK61422806, DK61422763, DK61523517, DK61422957, DK61523383, DK61523442, DK61322227, DK61422843, DK61523495, DK61423030, DK61422949, DK61322271, DK61322127, DK60621383, DK60621482, DK61322213, DK60321063, DK60320788, DK60621426, DK60621646, DK61322238, DK60621223, DK60621517, DK61322379, DK52316002, DK60320966, DK60621224, DK60621197, DK60821808, DK60621565, DK60821916, DK60821936, DK60320570, DK60320720, DK60320768, DK60320506, DK60320537, DK60321057, DK60320711, DK60320892, DK60320761, DK60320544, DK60321033, DK60320571, DK60320759, DK60320484, DK60321044, DK60320839, DK60320419, DK60320437,…
- FDA recall number
- Z-0199-2027
- Quantity
- 174
The text above is the beginning of the record; 660 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips North America (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.