What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
390 more codes are in the record below.
Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 1. 782185 (SE); UDI-DI: 00884838125698; Serial Number: 53002; 2. Model Number (REF): 782192 (XE); UDI-DI: 00884838128422; Serial Numbers: 56000, 56001, 56005, 56010, 56012;
- FDA recall number
- Z-3307-2026
- Quantity
- 6
Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 61047, 61056, 61057, 61076, 61087, 61088, 61105, 61116; 2. Model Number (REF): 781341 (1.5T); UDI-DI: 00884838055322; Serial Numbers: 70042, 70051, 70056, 70062, 70203, 70227, 70229, 70249, 70257, 70262, 70285, 70300, 70311, 70383, 70424, 70432, 70438, 70445, 70453, 70460, 70472, 70588, 70598, 70619, 70638, 70642, 70645, 70647, 70708, 70717, 70724, 70728, 70743, 70750, 70751, 70787, 70801, 70821, 70825, 70836, 70846, 70875, 70876, 70886, 70889, 70896, 70904, 70920, 70957, 70961, 70967, 70970, 70979, 70988, 84049, 84076, 84079, 84080, 84141, 84162, 84170, 84173, 84186, 84205, 84212, 84219, 84233, 84250, 84254, 84266, 84267, 84281, 84289, 84329, 84332, 84335, 84343, 84359, 84360, 84368, 84379, 84388, 84392, 84403, 84415, 84431, 84433, 84436, 84450, 84451, 84467, 84476, 84491, 84500, 84510, 84539, 84549, 84555, 84556, 84568, 84596, 84634, 84635, 84641, 84642, 84652, 84660, 84708, 84713, 84714, 84721, 84741, 84745, 84768, 84804, 84839, 84840, 84853, 84858, 84867, 84869, 84893, 84904, 84908, 84978, 84987, 87068, 87076, 87092, 87196, 87197, 87259; 3. Model Number (REF): 781396 (1.5T); UDI-DI: 00884838009820; Serial Numbers: 41000, 41001, 41021, 41034, 41046, 41056, 41061, 41091, 41116, 41145, 41194, 41231, 41249, 41253, 41254, 41255, 41352, 41354, 41356, 41370, 41400, 41408, 41422, 41428, 41456, 41475, 41492, 41513, 41515, 41548; 4. Model Number (REF): 782115 (1.5T);…
- FDA recall number
- Z-3308-2026
- Quantity
- 216
The text above is the beginning of the record; 566 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 1. 781342 (3.0T); UDI-DI: 00884838055339; Serial Numbers: 71001, 71002, 71008, 71031, 71038, 71051, 71061, 71072, 71084, 71088, 71089, 71095, 71101, 71154, 71163, 71164, 71166, 71184, 71185, 71193, 71210, 71231, 71269, 71271, 71288, 71290, 71297, 71301, 71303, 71309, 71311, 71312, 71322, 71335, 71336, 71340, 71341, 71344, 71355, 71375, 71385, 71393, 71412, 71413, 71420, 71422, 71427, 71438, 71441, 71443, 71445, 71446, 71449, 71450, 71459, 71462, 71463, 71475, 71486, 71487, 71489, 71491, 71513, 71521, 71528, 71556, 71558, 71562, 71589, 71604, 71607, 71612, 71615, 71619, 71620, 71639, 71641, 71646, 71665, 71673, 71688, 71697, 71704, 71710, 71715, 71729, 71730, 71739, 71754, 71756, 71757, 71759, 71813, 71817, 71828, 71831, 71841, 71849, 71850, 71853, 71862, 71885, 71896, 71921, 71923, 71954, 71959, 71960, 71962, 71963, 71964, 71986, 71992, 85006, 85020, 85032, 85038, 85046, 85066, 85068, 85076, 85078, 85085, 85087, 85091, 85092, 85102, 85126, 85148, 85150, 85154, 85191, 85201, 85216, 85245, 85279, 85377; 2. Model Number (REF): 781377 (3.0T); UDI-DI: 00884838009813; Serial Numbers: 42012, 42014, 42015, 42017, 42033, 42036, 42048, 42059, 42088, 42092, 42109, 42125, 42126, 42127, 42146, 42169, 42196, 42198, 42206, 42210, 42218, 42220, 42224, 42228, 42234, 42238, 42247, 42258, 42265, 42271, 42282, 42288, 42293, 42305, 42307, 42321, 42322, 42330, 42342, 42352, 42353, 42359, 42360, 42386, 42389, 42396, 42403, 42409, 42412, 42416, 42425; 3. Model Number (REF):…
- FDA recall number
- Z-3309-2026
- Quantity
- 243
The text above is the beginning of the record; 521 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Ingenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (X), 9. 782160 (X), 10. 782195 (X); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 781359 (S); UDI-DI: 00884838090057; Serial Numbers: 48010, 48023, 48028, 48047, 48122, 48160, 48173, 48180, 48196, 48286; 2. Model Number (REF): 782108 (S); UDI-DI: 00884838098343; Serial Numbers: 48507, 48510, 48530, 48548, 48554, 48576, 48577, 48580, 48586, 48607, 48616, 48620, 48653, 48686, 48688, 48726, 48746, 48778, 48818, 48831, 48847, 48872, 48878, 48888, 48893, 48896, 48897, 48917, 48922, 48931; 3. Model Number (REF): 782139 (S); UDI-DI: 00884838108639; Serial Numbers: 29016, 29017, 29018, 29020, 29026, 29029, 29031, 29041, 29054, 29063, 29069, 29071, 29076, 29078, 29079, 29086, 29104, 29115, 29118, 29144, 29145, 29150, 29160, 29164, 29169; 4. Model Number (REF): 782159 (S); UDI-DI: 00884838115705; Serial Numbers: 91007, 91027, 91031, 91038, 91043, 91051; 5. Model Number (REF): 782194 (S); UDI-DI: 00884838128477; Serial Numbers: 22401, 22410; 6. Model Number (REF): 781356 (X); UDI-DI: 00884838090040; Serial Numbers: 47004, 47006, 47011, 47012, 47013, 47014, 47018, 47020, 47026, 47027, 47028, 47029, 47037, 47038, 47039, 47043, 47048, 47051, 47057, 47059, 47063, 47067, 47068, 47070, 47092, 47094, 47098, 47099, 47177, 47187, 47188, 47189, 47195, 47197, 47199, 47203, 47206, 47211, 47215, 47221, 47222, 47233, 47237, 47242, 47248, 47253, 47260, 47268, 47270, 47271, 47274, 47276, 47278, 47279, 47281, 47282, 47289, 47290, 47292, 47294, 47297, 47303, 47305, 47313, 47314, 47322, 47344, 47347, 47348, 47351, 47353, 47356, 47371; 7. Model Number (REF):…
- FDA recall number
- Z-3310-2026
- Quantity
- 422
The text above is the beginning of the record; 2,210 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 781357 (S); UDI-DI: 00884838088108; Serial Numbers: 46002, 46012, 46013, 46017, 46018, 46020, 46021, 46023, 46025, 46027, 46029, 46036, 46039, 46041, 46042, 46044, 46053, 46054, 46082, 46086, 46092; 2. Model Number (REF): 782106 (S); UDI-DI: 00884838098329; Serial Numbers: 46203, 46204, 46207, 46208, 46209, 46219, 46220, 46225, 46235, 46243, 46249, 46258, 46269, 46270, 46286, 46302, 46315, 46325, 46375; 3. Model Number (REF): 782137 (S); UDI-DI: 00884838108615; Serial Numbers: 64001, 64002, 64003, 64004, 64005, 64014, 64015, 64029, 64032, 64039, 64040, 64041, 64042, 64049, 64050, 64051, 64052; 4. Model Number (REF): 782157 (S); UDI-DI: 00884838115682; Serial Numbers: 8900, 89003; 5. Model Number (REF): 781358 (X); UDI-DI: 00884838088115; Serial Numbers: 45008, 45009, 45010, 45011, 45024, 45026, 45028, 45029, 45032, 45033, 45034, 45035, 45036, 45037, 45040, 45041, 45046, 45048, 45049, 45054, 45055, 45056, 45060, 45062, 45066, 45070, 45072, 45073, 45075, 45076, 45078, 45086, 45087, 45089, 45090, 45091, 45093, 45094, 45096, 45097, 45098, 45099, 45100, 45105, 45108, 45112, 45113, 45115, 45116, 45119, 45120, 45127, 45129, 45132, 45136, 45138, 45140, 45146, 45147, 45148, 45152, 45154, 45156, 45157, 45160, 45162, 45168, 45174, 45181, 45186, 45187, 45189, 45190, 45191, 45194, 45205, 45215, 45231, 45235, 45237, 45243, 45245, 45246, 45250, 45252, 45264, 45272, 45277, 45278, 45290, 45296, 45316, 45317, 45319, 45323, 45328, 45333, 45335, 45340, 45342, 45374, 45379,…
- FDA recall number
- Z-3311-2026
- Quantity
- 340
The text above is the beginning of the record; 1,628 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
MR 5300; Model number (REF): 1. 782110, 2. 782152, 3. 782156, 4. 782196; Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 782110; UDI-DI: 00884838099364; Serial Numbers: 63007, 63027, 63064, 63076, 63097, 63123, 63132, 63137, 63148, 63150, 63160, 63236, 63237, 63238, 63258, 63300, 63302, 63360, 63370, 63394, 63397, 63398, 63445, 63523, 63576, 63599, 63636, 63696, 63735, 63766, 63767, 63774, 63806; 2. Model Number (REF): 782152; UDI-DI: 00884838112841; Serial Numbers: 69000, 69007, 69008, 69012, 69015, 69016, 69017, 69030, 69048, 69067, 69068, 69070, 69071, 69076, 69078, 69089, 69094, 69096, 69097, 69101, 69102, 69110, 69121, 69124, 69126, 69128, 69134, 69135, 69137, 69143, 69150, 69154, 69160, 69161, 69171, 69175, 69180, 69193, 69194, 69198, 69209, 69212, 69238, 69239, 69240, 69241, 69247, 69249, 69250, 69263, 69264, 69265, 69266, 69287, 69289, 69290, 69299, 69300, 69304, 69307, 69309, 69310, 69312, 69316, 69318, 69319, 69320, 69321, 69338, 69342, 69344, 69345, 69366, 69367; 3. Model Number (REF): 782156; UDI-DI: 00884838115675; Serial Numbers: 93000, 93007, 93008, 93018, 93023, 93026, 93028, 93037, 93054, 93073, 93084, 93100, 93101, 93102, 93103, 93104, 93105, 93106; 4. Model Number (REF): 782196; UDI-DI: 00884838128446; Serial Numbers: 54016, 54024, 54036, 54040, 54050, 54054;
- FDA recall number
- Z-3312-2026
- Quantity
- 131
MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 782120; UDI-DI: 00884838104112; Serial Numbers: 65000, 65001, 65002, 65003, 65004, 65005, 65006, 65008, 65009, 65011, 65012, 65013, 65014, 65015, 65016, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65038, 65039, 65040, 65041, 65042, 65043, 65049, 65050, 65051, 65056, 65059, 65060, 65061, 65063, 65064, 65065, 65066, 65067; 2. Model Number (REF): 782153; UDI-DI: 00884838112858; Serial Numbers: 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68013, 68014, 68016, 68017, 68018, 68019, 68020, 68021, 68022, 68023, 68024, 68025, 68026, 68028, 68029, 68030, 68031, 68032, 68033, 68034, 68036, 68037, 68041, 68042, 68043, 68047, 68049, 68050; 3. Model Number (REF): 782155; UDI-DI: 00884838115668; Serial Numbers: 94000, 94002, 94004, 94006, 94007, 94008, 94009, 94011, 94012, 94013, 94014, 94015, 94016, 94021, 94023, 94024; 4. Model Number (REF): 782199; UDI-DI: 00884838128439; Serial Numbers: 19300, 19303, 19304, 19309, 19312, 19313;
- FDA recall number
- Z-3313-2026
- Quantity
- 107
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782165 (1.5T), 5. 782145 (3.0T), 6. 782164 (3.0T), 7. 781270 (XR and 3.0T), 8. 782113 (XR and 3.0T), 9. 782129 (XR and 3.0T), 10. 782118 (Ingenia Elition X), 11. 782144 (Ingenia Elition X), 12. 782163 (Ingenia Elition X); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 781260 (1.5T); UDI-DI: 00884838095076; Serial Numbers: 5479, 5490, 8057, 8118, 8260, 8330, 8392, 8819, 8884, 8954, 10044, 10170, 10236, 11062, 13018, 13029, 13047, 18663, 18872, 20033, 20115, 20164, 20319, 20322, 20413, 21011, 21062, 21133, 21142, 21156, 21278, 21577, 21583, 21621, 21775, 21880, 21969, 22157, 22248, 32049, 32245, 32262, 32286, 32322, 32845, 39024; 2. Model Number (REF): 782112 (1.5T); UDI-DI: 00884838098886; Serial Numbers: 12002, 21205, 21658, 30031, 32378, 32519, 33338; 3. Model Number (REF): 782146 (1.5T); UDI-DI: 00884838108691; Serial Numbers: 8935, 13013, 21202, 30308, 32570, 32634, 32679, 32683, 32750, 33074, 33479, 33611, 33714, 39040; 4. Model Number (REF): 782165 (1.5T); UDI-DI: 00884838115767; Serial Numbers: 21763, 22170, 32402, 32409, 32971, 33379, 33387, 33571, 39075, 72160, 72190; 5. Model Number (REF): 782145 (3.0T); UDI-DI: 00884838108684; Serial Numbers: 17377, 17385, 17435, 24010, 24062, 34256, 34282, 38059, 38118, 38127, 38170, 38251, 38278, 38294, 38295, 38302, 38336, 73002, 74004, 74007, 74030; 6. Model Number (REF): 782164 (3.0T); UDI-DI: 00884838115750; Serial Numbers: 34038, 34254, 34268, 38021, 38046, 38061, 38086, 38090, 38223, 38246, 74020, 74033; 7. Model Number (REF): 781270 (XR and 3.0T); UDI-DI: 00884838095083; Serial Numbers: 17239, 17246, 17264, 17269, 17283, 17299, 17356, 17357, 17369, 17406, 17416, 17425, 17434, 17464, 17468, 17473, 17514, 17544, 17550, 17564, 17565, 17569, 17589, 17590, 23026, 23033,…
- FDA recall number
- Z-3314-2026
- Quantity
- 200
The text above is the beginning of the record; 887 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
SmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 782178; UDI-DI: 00884838122239; Serial Numbers: 42165, 42189, 71081; 2. Model Number (REF): 782207; UDI-DI: 00884838128576; Serial Numbers: 42055, 42287;
- FDA recall number
- Z-3315-2026
- Quantity
- 5
Upgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;
- Lot or code
- Model number (REF): 782127; UDI-DI: 00884838105591; Serial Numbers: 65007;
- FDA recall number
- Z-3316-2026
- Quantity
- 1
Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;
- Lot or code
- Model number (REF): 782130; UDI-DI: 00884838104402; Serial Numbers: 42002, 42030, 42037, 42066, 42104, 42151, 42163, 42164, 42211, 42219, 42248, 42376, 45001, 45007, 45043, 45044, 45052, 45179, 45236, 45263, 45552, 71060, 71262, 71279, 71406, 71532, 71565, 71580, 71616, 71907;
- FDA recall number
- Z-3317-2026
- Quantity
- 30
Evolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117 (3.0T), 6. 782143 (3.0T), 7. 782162 (3.0T); Software Versions: R11, R12 and R13;
- Lot or code
- 1. Model number (REF): 782116 (1.5T); UDI-DI: 00884838099722; Serial Numbers: 41042, 41235, 41287, 41295, 41367, 41505, 41568, 70068, 70093, 70995, 70996, 84370; 2. Model number (REF): 782148 (1.5T); UDI-DI: 00884838108714; Serial Numbers: 41011, 41031, 41110, 41244, 41277, 41334, 41427, 41446, 41464, 41523, 41534, 70425, 70454, 70473, 70622, 70630, 70644, 70769, 70803, 70966, 70983, 84102, 84409, 84560, 84569, 84630, 84825; 3. Model number (REF): 782166 (1.5T); UDI-DI: 00884838115774; Serial Numbers: 41005, 41161, 41485, 41499, 41572, 70155, 70206, 70826, 84048, 84270, 84541; 4. Model number (REF): 782201 (1.5T); UDI-DI: 00884838128552; Serial Numbers: 84018, 84648; 5. Model number (REF): 782117 (3.0T); UDI-DI: 00884838099739; Serial Numbers: 42023, 42024, 42078, 42375, 71348, 85681, 85705; 6. Model number (REF): 782143 (3.0T); UDI-DI: 00884838108660; Serial Numbers: 42069, 42225, 42254, 42318, 42366, 42415, 42440, 71389, 71460, 71595, 71815, 71836, 85649, 85651, 85654, 85682, 85683, 85688, 85690, 85695, 85697, 85699, 85701, 85714; 7. Model number (REF): 782162 (3.0T); UDI-DI: 00884838115736; Serial Numbers: 42184, 42369, 71074, 71403, 71433, 71574, 71740, 85145;
- FDA recall number
- Z-3318-2026
- Quantity
- 91
Distribution (FDA): Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips North America (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.