What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
380 more codes are in the record below.
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
- Lot or code
- 1. Model Number (REF): 781196. UDI: None. Serial Numbers: 21957. 2. Model Number (REF): 781296. UDI: (01)00884838004108(21)32039, (01)00884838004108(21)32095. Serial Numbers: 32039, 32095.
- FDA recall number
- Z-1951-2026
- Quantity
- 3 units
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
- Lot or code
- 1. Model Number (REF): 781178. UDI: None. Serial Numbers: 20088.
- FDA recall number
- Z-1952-2026
- Quantity
- 1 unit
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
- Lot or code
- 1. Model Number (REF): 781278. UDI: (01)00884838004139(21)38103. Serial Numbers: 38103.
- FDA recall number
- Z-1953-2026
- Quantity
- 1 unit
Philips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.
- Lot or code
- 1. Model Number (REF): 782116. UDI: (01)00884838099722(21)41527, (01)00884838099722(21)70949, (01)00884838099722(21)84741. Serial Numbers: 41527, 70949, 84741. 2. Model Number (REF): 782148. UDI: (01)00884838108714(21)84560, (01)00884838108714(21)41277, (01)00884838108714(21)70769, (01)00884838108714(21)70983. Serial Numbers: 84560, 41277, 70769, 70983. 3. Model Number (REF): 782166. UDI: (01)00884838115774(21)70042. Serial Numbers: 70042.
- FDA recall number
- Z-1954-2026
- Quantity
- 8 units
Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.
- Lot or code
- 1. Model Number (REF): 782143. UDI: (01)00884838108660(21)42218, (01)00884838108660(21)42225. Serial Numbers: 42218, 42225. 2. Model Number (REF): 782162. UDI: (01)00884838115736(21)71403. Serial Numbers: 71403.
- FDA recall number
- Z-1955-2026
- Quantity
- 3 units
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808
- Lot or code
- 1. Model Number (REF): 781315. UDI: (01)00884838099715(21)61067, (01)00884838099715(21)61058, (01)00884838099715(21)61056. Serial Numbers: 61067 61058, 61056. 2. Model Number (REF): 781341. UDI: (01)00884838055322(21)70229, (01)00884838055322(21)84667, (01)00884838055322(21)70311, (01)00884838055322(21)70668, (01)00884838055322(21)70845, (01)00884838055322(21)70851, (01)00884838055322(21)87151, (01)00884838055322(21)87112, (01)00884838055322(21)84322, (01)00884838055322(21)84323, (01)00884838055322(21)84228, (01)00884838055322(21)84789, (01)00884838055322(21)84261, (01)00884838055322(21)84500, (01)00884838055322(21)84556, (01)00884838055322(21)87187, (01)00884838055322(21)84840, (01)00884838055322(21)70685, (01)00884838055322(21)84661, (01)00884838055322(21)70787, (01)00884838055322(21)70063, (01)00884838055322(21)70621, (01)00884838055322(21)70743, (01)00884838055322(21)70911, (01)00884838055322(21)84049, (01)00884838055322(21)84088, (01)00884838055322(21)84147, (01)00884838055322(21)84254, (01)00884838055322(21)84283, (01)00884838055322(21)84403, (01)00884838055322(21)84433, (01)00884838055322(21)84460, (01)00884838055322(21)84462, (01)00884838055322(21)84467, (01)00884838055322(21)84475, (01)00884838055322(21)84734, (01)00884838055322(21)84538. Serial Numbers: 70229, 84667, 70311, 70668, 70845, 70851, 87151, 87112, 84322, 84323, 84228, 84789, 84261, 84500, 84556, 87187, 84840, 70685, 84661, 70787, 70063, 70621, 70743, 70911, 84049, 84088, 84147, 84254, 84283, 84403, 84433,…
- FDA recall number
- Z-1956-2026
- Quantity
- 54 units
The text above is the beginning of the record; 650 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
- Lot or code
- 1. Model Number (REF): 781262. UDI: (01)00884838068445(21)77144. Serial Numbers: 77144.
- FDA recall number
- Z-1957-2026
- Quantity
- 1 unit
Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808
- Lot or code
- 1. Model Number (REF): 781342. UDI: (01)00884838055339(21)71693, (01)00884838055339(21)71581, (01)00884838055339(21)71582, (01)00884838055339(21)71607, (01)00884838055339(21)85216, (01)00884838055339(21)71106, (01)00884838055339(21)71477, (01)00884838055339(21)71512, (01)00884838055339(21)71966, (01)00884838055339(21)71967, (01)00884838055339(21)85147, (01)00884838055339(21)85224, (01)00884838055339(21)85318, (01)00884838055339(21)71462, (01)00884838055339(21)71619, (01)00884838055339(21)85250, (01)00884838055339(21)71002, (01)00884838055339(21)71037, (01)00884838055339(21)71269, (01)00884838055339(21)71841, (01)00884838055339(21)71885, (01)00884838055339(21)71897, (01)00884838055339(21)85126, (01)00884838055339(21)85109, (01)00884838055339(21)71296, (01)00884838055339(21)71188, (01)00884838055339(21)71088, (01)00884838055339(21)71387, (01)00884838055339(21)71799, (01)00884838055339(21)71040, (01)00884838055339(21)71344, (01)00884838055339(21)71583, (01)00884838055339(21)71916, (01)00884838055339(21)71368, (01)00884838055339(21)71756, (01)00884838055339(21)71822, (01)00884838055339(21)71459, (01)00884838055339(21)71022, (01)00884838055339(21)71185, (01)00884838055339(21)71286, (01)00884838055339(21)71378, (01)00884838055339(21)71646, (01)00884838055339(21)71732, (01)00884838055339(21)71612, (01)00884838055339(21)71340, (01)00884838055339(21)71336. Serial Numbers: 71693, 71581, 71582, 71607, 85216, 71106, 71477, 71512, 71966, 71967, 85147, 85224, 85318, 71462, 71619, 85250,…
- FDA recall number
- Z-1958-2026
- Quantity
- 64 units
The text above is the beginning of the record; 962 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
- Lot or code
- 1. Model Number (REF): 781271. UDI: (01)00884838068452(21)78017, (01)00884838068452(21)78013, (01)00884838068452(21)78014, (01)00884838068452(21)78034, (01)00884838068452(21)78058, (01)00884838068452(21)78073, (01)00884838068452(21)78075, (01)00884838068452(21)78092, (01)00884838068452(21)78112, (01)00884838068452(21)78123, (01)00884838068452(21)78201, (01)00884838068452(21)78202, (01)00884838068452(21)78215, (01)00884838068452(21)78217, (01)00884838068452(21)78239, (01)00884838068452(21)78247, (01)00884838068452(21)78260,(01)00884838068452(21)78286, (01)00884838068452(21)78287, (01)00884838068452(21)78026. Serial Numbers: 78017, 78013, 78014, 78034, 78058, 78073, 78075, 78092, 78112, 78123, 78201, 78202, 78215, 78217, 78239, 78247, 78260, 78286, 78287, 78026. 2. Model Number (REF): 782105. UDI: (01)00884838098312(21)78532, (01)00884838098312(21)78576. Serial Numbers: 78532, 78576.
- FDA recall number
- Z-1959-2026
- Quantity
- 22 units
Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
- Lot or code
- 1. Model Number (REF): 781359. UDI: (01)00884838090057(21)48218, (01)00884838090057(21)48233. Serial Numbers: 48218, 48233. 2. Model Number (REF): 782108. UDI: (01)00884838098343(21)48652, (01)00884838098343(21)48551, (01)00884838098343(21)48669. Serial Numbers: 48652, 48551, 48669. 3. Model Number (REF): 782139. UDI: (01)00884838108639(21)29076, (01)00884838108639(21)29079. Serial Numbers: 29076, 29079.
- FDA recall number
- Z-1960-2026
- Quantity
- 7 units
Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.
- Lot or code
- 1. Model Number (REF): 781356. UDI: (01)00884838090040(21)47056, (01)00884838090040(21)47272, (01)00884838090040(21)47189, (01)00884838090040(21)47310, (01)00884838090040(21)47006, (01)00884838090040(21)47029, (01)00884838090040(21)47044, (01)00884838090040(21)47068, (01)00884838090040(21)47196, (01)00884838090040(21)47206, (01)00884838090040(21)47208, (01)00884838090040(21)47237, (01)00884838090040(21)47260, (01)00884838090040(21)47289, (01)00884838090040(21)47291, (01)00884838090040(21)47294, (01)00884838090040(21)47303, (01)00884838090040(21)47305, (01)00884838090040(21)47322, (01)00884838090040(21)47325, (01)00884838090040(21)47330, (01)00884838090040(21)47339, (01)00884838090040(21)47380, (01)00884838090040(21)47387. Serial Numbers: 47056, 47272, 47189, 47310, 47006, 47029, 47044, 47068, 47196, 47206, 47208, 47237, 47260, 47289, 47291, 47294, 47303, 47305, 47322, 47325, 47330, 47339, 47380, 47387. 2. Model Number (REF): 782109. UDI: (01)00884838098350(21)47869, (01)00884838098350(21)47565, (01)00884838098350(21)47644, (01)00884838098350(21)47881, (01)00884838098350(21)47509, (01)00884838098350(21)47919, (01)00884838098350(21)47634, (01)00884838098350(21)47681, (01)00884838098350(21)47706, (01)00884838098350(21)47907, (01)00884838098350(21)47867, (01)00884838098350(21)47550, (01)00884838098350(21)47804, (01)00884838098350(21)47549, (01)00884838098350(21)47635, (01)00884838098350(21)47788, (01)00884838098350(21)47617, (01)00884838098350(21)47566,…
- FDA recall number
- Z-1961-2026
- Quantity
- 86 units
The text above is the beginning of the record; 1,830 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.
- Lot or code
- 1. Model Number (REF): 781357. UDI: (01)00884838088108(21)46002. Serial Numbers: 46002. 2. Model Number (REF): 782106. UDI: (01)00884838098329(21)46243, (01)00884838098329(21)46292, (01)00884838098329(21)46317. Serial Numbers: 46243, 46292, 46317. 3. Model Number (REF): 782137. UDI: (01)00884838108615(21)64034, (01)00884838108615(21)64045. Serial Numbers: 64034, 64045.
- FDA recall number
- Z-1962-2026
- Quantity
- 6 units
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808
- Lot or code
- 1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)45160, (01)00884838088115(21)45438, (01)00884838088115(21)45034, (01)00884838088115(21)45049, (01)00884838088115(21)45066, (01)00884838088115(21)45090, (01)00884838088115(21)45141, (01)00884838088115(21)45149, (01)00884838088115(21)45170, (01)00884838088115(21)45250, (01)00884838088115(21)45290, (01)00884838088115(21)45085, (01)00884838088115(21)45312, (01)00884838088115(21)45376, (01)00884838088115(21)45122, (01)00884838088115(21)45169, (01)00884838088115(21)45191, (01)00884838088115(21)45192, (01)00884838088115(21)45203, (01)00884838088115(21)45204, (01)00884838088115(21)45207, (01)00884838088115(21)45220, (01)00884838088115(21)45228, (01)00884838088115(21)45229, (01)00884838088115(21)45249, (01)00884838088115(21)45278, (01)00884838088115(21)45279, (01)00884838088115(21)45320, (01)00884838088115(21)45028, (01)00884838088115(21)45195, (01)00884838088115(21)45174, (01)00884838088115(21)45165, (01)00884838088115(21)45382, (01)00884838088115(21)45011, (01)00884838088115(21)45055, (01)00884838088115(21)45075, (01)00884838088115(21)45113, (01)00884838088115(21)45147, (01)00884838088115(21)45152, (01)00884838088115(21)45235, (01)00884838088115(21)45237, (01)00884838088115(21)45245, (01)00884838088115(21)45264, (01)00884838088115(21)45269, (01)00884838088115(21)45277, (01)00884838088115(21)45316, (01)00884838088115(21)45322, (01)00884838088115(21)45337,…
- FDA recall number
- Z-1963-2026
- Quantity
- 103 units
The text above is the beginning of the record; 2,378 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
- Lot or code
- 1. Model Number (REF): 782120. UDI: (01)00884838104112(21)65019, (01)00884838104112(21)65004, (01)00884838104112(21)65028, (01)00884838104112(21)65031, (01)00884838104112(21)65039, (01)00884838104112(21)65056, (01)00884838104112(21)65066, (01)00884838104112(21)65065, (01)00884838104112(21)65038, (01)00884838104112(21)65005, (01)00884838104112(21)65043, (01)00884838104112(21)65049, (01)00884838104112(21)65051, (01)00884838104112(21)65060. Serial Numbers: 65019, 65004, 65028, 65031, 65039, 65056, 65066, 65065, 65038, 65005, 65043, 65049, 65051, 65060. 2. Model Number (REF): 782153. UDI: (01)00884838112858(21)68034, (01)00884838112858(21)68042, (01)00884838112858(21)68014, (01)00884838112858(21)68001, (01)00884838112858(21)68007, (01)00884838112858(21)68008, (01)00884838112858(21)68017, (01)00884838112858(21)68025, (01)00884838112858(21)68037, (01)00884838112858(21)68041, (01)00884838112858(21)68019. Serial Numbers: 68034, 68042, 68014, 68001, 68007, 68008, 68017, 68025, 68037, 68041, 68019.
- FDA recall number
- Z-1964-2026
- Quantity
- 25 untis
Philips SmartPath to dStream for 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781260. 2. Model Number (REF): 782112.
- Lot or code
- 1. Model Number (REF): 781260. UDI: (01)00884838095076(21)21554, (01)00884838095076(21)13018, (01)00884838095076(21)12033. Serial Numbers: 21554, 13018, 12033. 2. Model Number (REF): 782112. UDI: (01)00884838098886(21)30143, (01)00884838098886(21)22051, (01)00884838098886(21)32156. Serial Numbers: 30143, 22051, 32156.
- FDA recall number
- Z-1965-2026
- Quantity
- 6 units
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
- Lot or code
- 1. Model Number (REF): 782145. UDI: (01)00884838108684(21)17385. Serial Numbers: 17385.
- FDA recall number
- Z-1966-2026
- Quantity
- 1 unit
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
- Lot or code
- 1. Model Number (REF): 781270. UDI: (01)00884838095083(21)38109, (01)00884838095083(21)17386, (01)00884838095083(21)38356, (01)00884838095083(21)34211, (01)00884838095083(21)38113, (01)00884838095083(21)17356, (01)00884838095083(21)38111, (01)00884838095083(21)74011, (01)00884838095083(21)24058, (01)00884838095083(21)24066, (01)00884838095083(21)34223. Serial Numbers: 38109, 17386, 38356, 34211, 38113, 17356, 38111, 74011, 24058, 24066, 34223. 2. Model Number (REF): 782113. UDI: (01)00884838098909(21)38066, (01)00884838098909(21)38117, (01)00884838098909(21)17384. Serial Numbers: 38066, 38117, 17384.
- FDA recall number
- Z-1967-2026
- Quantity
- 14 units
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
- Lot or code
- 1. Model Number (REF): 782118. UDI: (01)00884838099746(21)85655. Serial Numbers: 85655. 2. Model Number (REF): 782144. UDI: (01)00884838108677(21)71087. Serial Numbers: 71087. 3. Model Number (REF): 782163. UDI: (01)00884838115743(21)42299. Serial Numbers: 42299.
- FDA recall number
- Z-1968-2026
- Quantity
- 3 units
Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
- Lot or code
- 1. Model Number (REF): 782130. UDI: (01)00884838104402(21)45052, (01)00884838104402(21)42030, (01)00884838104402(21)42151, (01)00884838104402(21)71279, (01)00884838104402(21)71565, (01)00884838104402(21)42104, (01)00884838104402(21)42163, (01)00884838104402(21)42376, (01)00884838104402(21)45179, (01)00884838104402(21)45236, (01)00884838104402(21)45263, (01)00884838104402(21)71532, (01)00884838104402(21)85394, (01)00884838104402(21)85643, (01)00884838104402(21)85644. Serial Numbers: 45052, 42030, 42151, 71279, 71565, 42104, 42163, 42376, 45179, 45236, 45263, 71532, 85394, 85643, 85644.
- FDA recall number
- Z-1969-2026
- Quantity
- 15 units
Distribution (FDA): Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.