What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
31 more codes are in the record below.
Achieva XR Product Numbers: (1) 781153, (2) 781253;
- Lot or code
- Product Numbers: (1) 781153, (2) 781253; UDI-DI: (1) no UDI, (2) no UDI; Serial Numbers: All serial numbers;
- FDA recall number
- Z-1928-2025
- Quantity
- 6 units (5 US, 1 OUS)
Intera 1.5T Achieva IT Nova Product Number: 781175
- Lot or code
- Product Number: 781175; UDI-DI: no UDI; Serial Number: 15003;
- FDA recall number
- Z-1929-2025
- Quantity
- 1 unit
Intera 1.5T Achieva Nova Product Number: 781172;
- Lot or code
- Product Number: 781172; UDI-DI: no UDI; Serial Numbers: 21218, 21217, 21216, 21289, 21349, 12146, 12063, 12086, 20069, 21027, 21198, 12099, 18641, 21476, 18623, 12078;
- FDA recall number
- Z-1930-2025
- Quantity
- 16 units (1 US, 15 OUS)
Intera 1.5T Achieva Nova-Dual Product Number: 781173;
- Lot or code
- Product Number: 781173; UDI-DI: no UDI; Serial Numbers: 13010, 13006, 21501;
- FDA recall number
- Z-1931-2025
- Quantity
- 3 units (1 US, 2 OUS)
Intera 1.5T Explorer/Nova Dual Product Number: 781108;
- Lot or code
- Product Number: 781108; UDI-DI: no UDI; Serial Number: 9025;
- FDA recall number
- Z-1932-2025
- Quantity
- 1 unit (OUS only)
Intera 1.5T Master/Nova Product Number: 781106;
- Lot or code
- Product Number: 781106; UDI-DI: no UDI; Serial Numbers: 8009, 18863;
- FDA recall number
- Z-1933-2025
- Quantity
- 2 units (US only)
Intera 1.5T Omni/Stellar; Product Number: 781104;
- Lot or code
- Product Number: 781104; UDI-DI: no UDI; Serial Numbers: 10403, 18969;
- FDA recall number
- Z-1934-2025
- Quantity
- 2 units (OUS only)
Intera Achieva 1.5T Pulsar; Product Number: 781171;
- Lot or code
- Product Number: 781171; UDI-DI: no UDI; Serial Numbers: 11044, 11039, 18627, 11069;
- FDA recall number
- Z-1935-2025
- Quantity
- 4 units (1 US, 3 OUS)
SmartPath to dStream for 1.5T¿; Product Number: 782146;
- Lot or code
- Product Number: 782146; UDI-DI: no UDI; Serial Number: 13013;
- FDA recall number
- Z-1936-2025
- Quantity
- 1 unit
Distribution (FDA): Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips North America (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.