GE Medical Systems, LLC recalls
21 notices naming GE Medical Systems, LLC as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
GE Medical Systems, LLC: Senographe Pristina and 2 more
Hazard (FDA): GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is…
FDAMedical devices
GE Medical Systems, LLC: Optima XR200amx and 2 more
Hazard (FDA): GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly…
FDAMedical devices
GE Medical Systems, LLC: Giraffe OmniBed and 2 more
Hazard (FDA): GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners…
FDAMedical devices
GE Medical Systems, LLC: Allia IGS 3 Pulse angiographic X-ray system and 4 more
Hazard (FDA): GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel…
FDAMedical devices
GE Medical Systems, LLC: Revolution Ascend and 6 more
Hazard (FDA): GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution…
FDAMedical devices
GE Medical Systems, LLC: Imactis CT-Navigation System
Hazard (FDA): There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
FDAMedical devices
GE Medical Systems, LLC: Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1 and 2 more
Hazard (FDA): There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized…
FDAMedical devices
GE Medical Systems, LLC: Senographe Pristina
Hazard (FDA): Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR ¿1020.30(j).
FDAMedical devices
GE Medical Systems, LLC: MAC VU360 Acquisition Trunk Cable and Module Holder
Hazard (FDA): If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires…
FDAMedical devices
GE Medical Systems, LLC: LOGIQ P10 series with software version R4.5.7 Model Number 5877534 and 2 more
Hazard (FDA): The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
FDAMedical devices
GE Medical Systems, LLC: GE HealthCare OEC 9800. and 1 more
Hazard (FDA): GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator…
FDAMedical devices
GE Medical Systems, LLC: GE Signa Artist and 21 more
Hazard (FDA): GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T…
FDAMedical devices
GE Medical Systems, LLC: SIGNA Architect and 3 more
Hazard (FDA): GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result…
FDAMedical devices
GE Medical Systems, LLC: OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital… and 1 more
Hazard (FDA): OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
FDAMedical devices
Fire
GE Medical Systems, LLC: Venue Go Models: R2 and 1 more
Hazard (FDA): The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
FDAMedical devices
GE Medical Systems, LLC: GE Healthcare Revolution Apex and 6 more
Hazard (FDA): GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol)…
FDAMedical devices
GE Medical Systems, LLC: Definium 656HD & Discovery XR656HD and 3 more
Hazard (FDA): The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still…
FDAMedical devices
GE Medical Systems, LLC: Portrait Core Services
Hazard (FDA): There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
FDAMedical devices
GE Medical Systems, LLC: GE Healthcare Optima XR200amx and 2 more
Hazard (FDA): GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and…
FDAMedical devices
GE Medical Systems, LLC: Tec 6 Plus Anesthesia Vaporizer and 4 more
Hazard (FDA): Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
FDAMedical devices
GE Medical Systems, LLC: Discovery MR750w 3.0T and 3 more
Hazard (FDA): Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
FDAMedical devices