What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
- Lot or code
- GTINs: 00840682138598, 00195278840226, 00840682136709)
- FDA recall number
- Z-1043-2025
- Quantity
- 3149 (1697 US; 1452 OUS)
Discovery XR656HD. X-Ray imaging system.
- Lot or code
- GTINs: 00840682142175, 00840682142168
- FDA recall number
- Z-1044-2025
- Quantity
- 3149 (1697 US; 1452 OUS)
Discovery XR656 HD. X-Ray imaging system.
- Lot or code
- Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265)
- FDA recall number
- Z-1045-2025
- Quantity
- 3149 (1697 US; 1452 OUS)
Optima XR646 HD. X-Ray imaging system.
- Lot or code
- GTIN: 00195278526892
- FDA recall number
- Z-1046-2025
- Quantity
- 3149 (1697 US; 1452 OUS)
Distribution (FDA): Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE Medical Systems, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.