What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Optima XR200amx, Mobile Digital-Ready Radiographic System
- Lot or code
- UDI-DI: 00840682115278; System IDs: 419383DP3, 773792XR200, 856355OPT, FEB520100
- FDA recall number
- Z-2726-2026
- Quantity
- 4 units
Optima XR220amx, Mobile Digital-Ready Radiographic System
- Lot or code
- UDI-DI: 00840682115261; System IDs: 05249MOB02.OLD, 205820DP1, 228818XR220, 229276M1, 240566P1, 301295OPTIMA2, 303788SBOPT1, 314251P5, 352LEEDP1, 386FLGDPXR3, 417347DPORT1, 512SAMCSB, 516632OPTIMA2, 540981ODP7, 563264OPT, 601445DP1, 615338HDP01, 661949DP2, 713441OPCDX1, 717657OPT1, 718206OPT240A, 727848SB1, 732321DP1, 757736DP3, 813302OPT1, 828213DXP1, 828286OPT1, 843847220D, 850070941, 865545DP3, 915821OPT1, 941342OP221, 972566XR220E, 985447TM4, SA1208MM05, SB2201700085WK, XR1034798WK7, XR1037005WK4
- FDA recall number
- Z-2727-2026
- Quantity
- 38 units
Optima XR240amx, Mobile Digital Radiographic System
- Lot or code
- UDI-DI: 00840682122351 and 00840682123686; System IDs: 252847PP249, 281MSL2402, 291OPTI1, 312CCHSUPER7, 318487OPTIMA1, 360744OPT2402, 501686OPTI1, 504826XR241, 519941XR240A, 540776ODP3, 541472M4, 602344240C, 706291BEE9, 713799XR240, 770836BEE2, 813872OPT11, 813872OPT9, 843724OPT1, 843847OPT240, 870239OPT2, 905522AMX240, 973625DIG4, DF2402001002WK, SA1013MM13
- FDA recall number
- Z-2728-2026
- Quantity
- 24 units
Distribution (FDA): Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE Medical Systems, LLC
- Title