Why it was recalled
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
306 more codes are in the record below.
GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number: 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number: 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number: 630266CTRV, UDI (01)00195278460271(11)240800(21)REV2X2400093CN, Mfg. Lot/Serial Number REV2X2400093CN; System ID Number: 773702REVCT6ED, UDI (01)00195278460271(11)231000(21)REV2X2300071CN, Mfg. Lot/Serial Number REV2X2300071CN; System ID Number: 312996CTRV2, UDI (01)00195278460271(11)241000(21)REV2X2400126CN, Mfg. Lot/Serial Number REV2X2400126CN; System ID Number: 1000000018, UDI To be provided, Mfg. Lot/Serial Number REV2X2300074CN; System ID Number: 617726Y6REV2, UDI (01)00840682118552(11)170900(21)REVVX1700090CN, Mfg. Lot/Serial Number REVVX1700090CN; System ID Number: 617726Y6REV1, UDI (01)00840682118552(11)160700(21)REVVX1600018CN, Mfg. Lot/Serial Number REVVX1600018CN; System ID Number: 336890APEX, UDI (01)00195278460271(11)240900(21)REV2X2400116CN, Mfg. Lot/Serial Number REV2X2400116CN; System ID Number: 704801APEX, UDI (01)00195278460271(11)241000(21)REV2X2400139CN, Mfg. Lot/Serial Number REV2X2400139CN; System ID Number: 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number: 980993CT1, UDI (01)00840682146616(11)210700(21)REV2A2100029CN, Mfg. Lot/Serial Number REV2A2100029CN; System ID Number: 910715REV1, UDI (01)00195278460271(11)241200(21)REV2X2400155CN, Mfg. Lot/Serial…
- FDA recall number
- Z-1297-2025
- Quantity
- 56 units
The text above is the beginning of the record; 4,857 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 256265STCT, UDI (01)00195278460271(11)230600(21)REV2X2300021CN, Mfg. Lot/Serial Number REV2X2300021CN; System ID Number 480342APEX1, UDI (01)00195278460271(11)240600(21)REV2X2400065CN, Mfg. Lot/Serial Number REV2X2400065CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858824REV, UDI (01)00195278460271(11)241100(21)REV2X2400147CN, Mfg. Lot/Serial Number REV2X2400147CN; System ID Number 707257APEX1, UDI (01)00195278460271(11)231000(21)REV2X2300077CN, Mfg. Lot/Serial Number REV2X2300077CN; System ID Number 714456APEX, UDI (01)00195278460271(11)231200(21)REV2X2300094CN, Mfg. Lot/Serial Number REV2X2300094CN; System ID Number 719227APEX1, UDI (01)00195278460271(11)240200(21)REV2X2400019CN, Mfg. Lot/Serial Number REV2X2400019CN; System ID Number 970945CT, UDI (01)00195278460271(11)240400(21)REV2X2400046CN, Mfg. Lot/Serial Number REV2X2400046CN; System ID Number 203576EDAPEX, UDI (01)00195278460271(11)240300(21)REV2X2400021CN, Mfg. Lot/Serial Number REV2X2400021CN; System ID Number 860224EDAPEX, UDI (01)00195278460271(11)240500(21)REV2X2400050CN, Mfg. Lot/Serial Number REV2X2400050CN; System ID Number 203276APEX1, UDI…
- FDA recall number
- Z-1298-2025
- Quantity
- 250 units
The text above is the beginning of the record; 27,085 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 205608CT, UDI (01)00840682123471(11)180800(21)REVV81800010CN, Mfg. Lot/Serial Number REVV81800010CN; System ID Number 256265STCT, UDI (01)00195278460271(11)230600(21)REV2X2300021CN, Mfg. Lot/Serial Number REV2X2300021CN; System ID Number 205699CTREV, UDI Not Applicable, Mfg. Lot/Serial Number 00000442592CN2; System ID Number 480342APEX, UDI (01)00840682146616(11)220900(21)REV2A2200072CN, Mfg. Lot/Serial Number REV2A2200072CN; System ID Number 480342APEX1, UDI (01)00195278460271(11)240600(21)REV2X2400065CN, Mfg. Lot/Serial Number REV2X2400065CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 11320, UDI To be provided, Mfg. Lot/Serial Number REVVX2000033CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858657APEX, UDI (01)00840682118552(11)160900(21)REVVX1600034CN, Mfg. Lot/Serial Number REVVX1600034CN; System ID Number 858824REV, UDI (01)00195278460271(11)241100(21)REV2X2400147CN, Mfg. Lot/Serial Number REV2X2400147CN; System ID Number 858824APEX, UDI Not Applicable, Mfg. Lot/Serial Number 00000443327CN2; System ID Number 310423CT, UDI (01)00840682118552(11)211200(21)REVVX2100047CN, Mfg. Lot/Serial Number REVVX2100047CN; System ID Number 707257APEX1, UDI…
- FDA recall number
- Z-1299-2025
- Quantity
- 42 units
The text above is the beginning of the record; 3,331 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number 973971CTREV, UDI (01)00840682118552(11)160600(21)REVVX1600010CN, Mfg. Lot/Serial Number REVVX1600010CN; System ID Number 513584APEX1, UDI (01)00840682146616(11)230300(21)REV2A2300023CN, Mfg. Lot/Serial Number REV2A2300023CN
- FDA recall number
- Z-1300-2025
- Quantity
- 4 units
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN
- FDA recall number
- Z-1301-2025
- Quantity
- 1 unit
GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 480342APEX, UDI (01)00840682146616(11)220900(21)REV2A2200072CN, Mfg. Lot/Serial Number REV2A2200072CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858657APEX, UDI (01)00840682118552(11)160900(21)REVVX1600034CN, Mfg. Lot/Serial Number REVVX1600034CN; System ID Number 310423CT, UDI (01)00840682118552(11)211200(21)REVVX2100047CN, Mfg. Lot/Serial Number REVVX2100047CN; System ID Number 707257APEX1, UDI (01)00195278460271(11)231000(21)REV2X2300077CN, Mfg. Lot/Serial Number REV2X2300077CN; System ID Number 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number 408559CTREVO, UDI To be provided, Mfg. Lot/Serial Number REVV81900011CN; System ID Number 904UFNAPEX, UDI (01)00195278460271(11)230300(21)REV2X2300001CN, Mfg. Lot/Serial Number REV2X2300001CN; System ID Number 813972APEX2, UDI (01)00195278460271(11)240700(21)REV2X2400088CN, Mfg. Lot/Serial Number REV2X2400088CN; System ID Number 813972APEX3, UDI (01)00195278460271(11)241000(21)REV2X2400135CN, Mfg. Lot/Serial Number REV2X2400135CN; System ID Number 808522RCT, UDI Not Applicable, Mfg. Lot/Serial Number 00000441716CN8; System ID Number 312996CTRV2,…
- FDA recall number
- Z-1302-2025
- Quantity
- 105 units
The text above is the beginning of the record; 10,430 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
- Lot or code
- System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 650493REV2, UDI To be provided, Mfg. Lot/Serial Number REVVX1900036CN; System ID Number 864512APEX1, UDI (01)00195278460271(11)240500(21)REV2X2400059CN, Mfg. Lot/Serial Number REV2X2400059CN; System ID Number 832325REV, UDI (01)00195278460271(11)241000(21)REV2X2400134CN, Mfg. Lot/Serial Number REV2X2400134CN; System ID Number 281MWREVCT, UDI To be provided, Mfg. Lot/Serial Number REVVX1900008CN; System ID Number 713792ACT39, UDI To be provided, Mfg. Lot/Serial Number REVV82100006CN; System ID Number 206320REV, UDI To be provided, Mfg. Lot/Serial Number 00000441459CN5; System ID Number 206320REV2, UDI To be provided, Mfg. Lot/Serial Number REVV82200025CN; System ID Number 082421190701, UDI (01)00195278633149(11)241100(21)CBFCG2400026HM, Mfg. Lot/Serial Number CBFCG2400026HM; System ID Number DK1044CT12, UDI (01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/Serial Number REV2A2000001CN; System ID Number DK1547CT02, UDI (01)00840682146616(11)210100(21)REV2A2100002CN, Mfg. Lot/Serial Number REV2A2100002CN; System ID Number 11525316, UDI To be provided, Mfg. Lot/Serial Number REV2X2400082CN; System ID Number 259056CT01, UDI (01)00195278460271(11)230800(21)REV2X2300051CN, Mfg. Lot/Serial Number REV2X2300051CN; System ID Number 11437190, UDI To be provided, Mfg. Lot/Serial Number CBGLG2400001HM; System ID Number 0850211094, UDI…
- FDA recall number
- Z-1303-2025
- Quantity
- 21 units
The text above is the beginning of the record; 731 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide distribution.