What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
GE HealthCare OEC 9800.
- Lot or code
- GTIN: 00840682114349
- FDA recall number
- Z-2341-2025
GE HealthCare OEC 9900 Elite C-arm Systems.
- Lot or code
- GTIN: 00840682114349
- FDA recall number
- Z-2342-2025
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE Medical Systems, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.