What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
SIGNA Architect
- Lot or code
- GTIN: 00840682147095 00840682122702 00195278023643 00840682123440
- FDA recall number
- Z-2130-2025
SIGNA Architect AIR
- Lot or code
- GTIN: 00195278283481
- FDA recall number
- Z-2131-2025
Discovery MR750w 3.0T
- Lot or code
- GTIN: 00840682103817 00195278229519
- FDA recall number
- Z-2132-2025
SIGNA PET/MR
- Lot or code
- GTIN: 00840682105378
- FDA recall number
- Z-2133-2025
Distribution (FDA): Worldwide distribution - US Nationwide and the country of China.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE Medical Systems, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.