Siemens Medical Solutions USA, Inc recalls
15 notices naming Siemens Medical Solutions USA, Inc as the recalling firm, since February 5, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Siemens Medical Solutions USA, Inc: ARTIS icono ceiling. Model Number: 11328100.
Hazard (FDA): Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
FDAMedical devices
Siemens Medical Solutions USA, Inc: ARTIS icono biplane. Model Number: 11327600. and 2 more
Hazard (FDA): Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
FDAMedical devices
Siemens Medical Solutions USA, Inc: ARTIS pheno and 1 more
Hazard (FDA): During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the…
FDAMedical devices
Siemens Medical Solutions USA, Inc: ARTIS Pheno VE30A and VE40A
Hazard (FDA): During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the…
FDAMedical devices
Siemens Medical Solutions USA, Inc: ARTIS pheno and ARTIS Icono biplane
Hazard (FDA): During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels…
FDAMedical devices
Siemens Medical Solutions USA, Inc: SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel… and 3 more
Hazard (FDA): To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
FDAMedical devices
Siemens Medical Solutions USA, Inc: Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R and 1 more
Hazard (FDA): Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
FDAMedical devices
Siemens Medical Solutions USA, Inc: LUMINOS Lotus Max
Hazard (FDA): A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ…
FDAMedical devices
Siemens Medical Solutions USA, Inc: MAMMOMAT Revelation and 2 more
Hazard (FDA): There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
FDAMedical devices
Siemens Medical Solutions USA, Inc: MAGNETOM Cima.X Upgrade. Model Number: 11689304.
Hazard (FDA): There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting…
FDAMedical devices
Siemens Medical Solutions USA, Inc: Biograph mMR. Model Number: 10433372. and 14 more
Hazard (FDA): There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a…
FDAMedical devices
Siemens Medical Solutions USA, Inc: Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
Hazard (FDA): Limited system movements after startup .
FDAMedical devices
Siemens Medical Solutions USA, Inc: ARTIS is a family of dedicated angiography systems developed for single and biplane… and 3 more
Hazard (FDA): In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high…
FDAMedical devices
Fire
Siemens Medical Solutions USA, Inc: UROSKOP Omnia. Model Number: 10094910 and 1 more
Hazard (FDA): A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
FDAMedical devices
SIEMENS MEDICAL SOLUTIONS USA, INC: ARTIS One Angiographic X-Ray System
Hazard (FDA): A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue…
FDAMedical devices