FDA recall 97530: Siemens Medical Solutions USA, Inc: Biograph mMR. Model Number: 10433372. and 14 more

FDA recall 97530 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

384 more codes are in the record below.

  • Biograph mMR. Model Number: 10433372.

    Lot or code
    Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007.
    FDA recall number
    Z-2654-2025
    Quantity
    27 units
  • BIOGRAPH One (DE). Model Number: 11689172.

    Lot or code
    Model Number: 11689172. UDI Numbers: (01)4068151020472(21)238008. Serial Number: 238008.
    FDA recall number
    Z-2655-2025
    Quantity
    1 unit
  • MAGNETOM Cima.X (DE). Model Number: 11647158.

    Lot or code
    Model Number: 11647158. UDI Numbers: (01)04056869975221(21)213029, (01)04056869975221(21)213006, (01)04056869975221(21)213062, (01)04056869975221(21)213043, (01)04056869975221(21)213041, (01)04056869975221(21)213034, (01)04056869975221(21)213035, (01)04056869975221(21)213028, (01)04056869975221(21)213031, (01)04056869975221(21)213018, (01)04056869975221(21)213061, (01)04056869975221(21)213005, (01)04056869975221(21)213021, (01)04056869975221(21)213022, (01)04056869975221(21)213056, (01)04056869975221(21)213030, (01)04056869975221(21)213039, (01)04056869975221(21)213025, (01)04056869975221(21)213063, (01)04056869975221(21)213057, (01)04056869975221(21)213026, (01)04056869975221(21)213020. Serial Numbers: 213029, 213006, 213062, 213043, 213041, 213034, 213035, 213028, 213031, 213018, 213061, 213005, 213021, 213022, 213056, 213030, 213039, 213025, 213063, 213057, 213026, 213020.
    FDA recall number
    Z-2656-2025
    Quantity
    22 units
  • MAGNETOM Connectom.X. Model Number: 11371480.

    Lot or code
    Model Number: 11371480. UDI Numbers: N/A. Serial Numbers: 237551.
    FDA recall number
    Z-2657-2025
    Quantity
    1 unit
  • MAGNETOM Lumina (DE). Model Number: 11344916.

    Lot or code
    Model Number: 11344916. UDI Numbers: (01)04056869230740(21)196458, (01)04056869230740(21)196591, (01)04056869230740(21)196655, (01)04056869230740(21)196355, (01)04056869230740(21)196733, (01)04056869230740(21)196668, (01)04056869230740(21)196477, (01)04056869230740(21)196343, (01)04056869230740(21)196553, (01)04056869230740(21)196519, (01)04056869230740(21)196366, (01)04056869230740(21)196782, (01)04056869230740(21)196617, (01)04056869230740(21)196684, (01)04056869230740(21)196502, (01)04056869230740(21)196539, (01)04056869230740(21)196280, (01)04056869230740(21)196626, (01)04056869230740(21)196511, (01)04056869230740(21)196643, (01)04056869230740(21)196796, (01)04056869230740(21)196551, (01)04056869230740(21)196495, (01)04056869230740(21)196741, (01)04056869230740(21)196621, (01)04056869230740(21)196611, (01)04056869230740(21)196544, (01)04056869230740(21)196724, (01)04056869230740(21)196776, (01)04056869230740(21)196752, (01)04056869230740(21)196372, (01)04056869230740(21)196350, (01)04056869230740(21)196841, (01)04056869230740(21)196252, (01)04056869230740(21)196508, (01)04056869230740(21)196332, (01)04056869230740(21)196496, (01)04056869230740(21)196753, (01)04056869230740(21)196375, (01)04056869230740(21)196738, (01)04056869230740(21)196260, (01)04056869230740(21)196470, (01)04056869230740(21)196765, (01)04056869230740(21)196334, (01)04056869230740(21)196288, (01)04056869230740(21)196409, (01)04056869230740(21)196737, (01)04056869230740(21)196286,…
    FDA recall number
    Z-2658-2025
    Quantity
    130 units

    The text above is the beginning of the record; 3,516 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Prisma. Model Number: 10849582.

    Lot or code
    Model Number: 10849582. UDI Numbers: (01)04056869006727(21)166324, (01)04056869006727(21)66104, (01)04056869006727(21)166097, (01)04056869006727(21)166024, (01)04056869006727(21)166018, (01)04056869006727(21)66109, (01)04056869006727(21)66111, (01)04056869006727(21)166114, (01)04056869006727(21)166242, (01)04056869006727(21)166268, (01)04056869006727(21)166200, (01)04056869006727(21)166047, (01)04056869006727(21)66047, (01)04056869006727(21)166292, (01)04056869006727(21)166041, (01)04056869006727(21)166007, (01)04056869006727(21)166037, (01)04056869006727(21)66112, (01)04056869006727(21)166025, (01)04056869006727(21)166003, (01)04056869006727(21)66102, (01)04056869006727(21)166254, (01)04056869006727(21)166074, (01)04056869006727(21)166281, (01)04056869006727(21)166220, (01)04056869006727(21)66027, (01)04056869006727(21)166221, (01)04056869006727(21)166321, (01)04056869006727(21)166193, (01)04056869006727(21)166012, (01)04056869006727(21)166121, (01)04056869006727(21)66053, (01)04056869006727(21)166160, (01)04056869006727(21)66094, (01)04056869006727(21)166310, (01)04056869006727(21)66028, (01)04056869006727(21)166102, (01)04056869006727(21)166034, (01)04056869006727(21)166030, (01)04056869006727(21)66106, (01)04056869006727(21)166235, (01)04056869006727(21)166273, (01)04056869006727(21)166201, (01)04056869006727(21)166345, (01)04056869006727(21)166225, (01)04056869006727(21)66044, (01)04056869006727(21)66018, (01)04056869006727(21)166006, (01)04056869006727(21)166009,…
    FDA recall number
    Z-2659-2025
    Quantity
    101 units

    The text above is the beginning of the record; 2,353 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.

    Lot or code
    Model Number: 10432915. UDI Numbers: (01)04056869006703(21)45046, (01)04056869006703(21)45115, (01)04056869006703(21)45398, (01)04056869006703(21)145760, (01)04056869006703(21)145811, (01)04056869006703(21)45549, (01)04056869006703(21)45834, (01)04056869006703(21)146099, (01)04056869006703(21)146110, (01)04056869006703(21)45047, (01)04056869006703(21)146093, (01)04056869006703(21)46066, (01)04056869006703(21)45788, (01)04056869006703(21)146082, (01)04056869006703(21)45124, (01)04056869006703(21)45185, (01)04056869006703(21)45332, (01)04056869006703(21)45466, (01)04056869006703(21)145390, (01)04056869006703(21)45732, (01)04056869006703(21)45918, (01)04056869006703(21)45380, (01)04056869006703(21)46182, (01)04056869006703(21)45735, (01)04056869006703(21)45233, (01)04056869006680(21)162026, (01)04056869006703(21)145008, (01)04056869006703(21)145473, (01)04056869006703(21)145095, (01)04056869006703(21)145664, (01)04056869006703(21)45200, (01)04056869006703(21)146098, (01)04056869006703(21)146086, (01)04056869006703(21)45805, (01)04056869006703(21)45812, (01)04056869006703(21)45397, (01)04056869006703(21)46264, (01)04056869006703(21)45499, (01)04056869006703(21)45749, (01)04056869006703(21)45289, (01)04056869006703(21)46220, (01)04056869006703(21)45369, (01)04056869006703(21)45267, (01)04056869006703(21)45253, (01)04056869006703(21)45060, (01)04056869006703(21)146103, (01)04056869006703(21)45333, (01)04056869006703(21)146123, (01)04056869006703(21)46147,…
    FDA recall number
    Z-2660-2025
    Quantity
    586 units

    The text above is the beginning of the record; 20,033 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Skyra fit. Model Number: 10849580.

    Lot or code
    Model Number: 10849580. UDI Numbers: (01)04056869006710(21)170068, (01)04056869006710(21)170053, (01)04056869006710(21)70022, (01)04056869006710(21)170049, (01)04056869006710(21)170018, (01)04056869006680(21)162078, (01)04056869006710(21)170109, (01)04056869006710(21)170039, (01)04056869006710(21)170066, (01)04056869006710(21)170023, (01)04056869006710(21)170030, (01)04056869006710(21)170004, (01)04056869006710(21)170016, (01)04056869006710(21)170077, (01)04056869006710(21)170058, (01)04056869006710(21)170002, (01)04056869006710(21)170050, (01)04056869006710(21)170010, (01)04056869006710(21)170019, (01)04056869006710(21)70026, (01)04056869006710(21)170056, (01)04056869006710(21)170061, (01)04056869006710(21)170105, (01)04056869006710(21)170017, (01)04056869006710(21)170041, (01)04056869006710(21)70032, (01)04056869006710(21)170111, (01)04056869006710(21)170048, (01)04056869006710(21)170101, (01)04056869006710(21)170100, (01)04056869006710(21)170060, (01)04056869006710(21)170099, (01)04056869006710(21)170036, (01)04056869006710(21)170007, (01)04056869006710(21)170001, (01)04056869006710(21)170009, (01)04056869006710(21)70020, (01)04056869006710(21)170112, (01)04056869006710(21)170113, (01)04056869006710(21)170054, (01)04056869006710(21)170047, (01)04056869006710(21)170046, (01)04056869006710(21)70023, (01)04056869006710(21)70027, (01)04056869006710(21)170110, (01)04056869006710(21)170095, (01)04056869006710(21)170075, (01)04056869006710(21)170011, (01)04056869006710(21)170081,…
    FDA recall number
    Z-2661-2025
    Quantity
    56 units

    The text above is the beginning of the record; 688 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.

    Lot or code
    Model Number: 11516217. UDI Numbers: (01)04056869299273(21)212087, (01)04056869299273(21)212098, (01)04056869299273(21)212088, (01)04056869299273(21)212068, (01)04056869299273(21)212081, (01)04056869299273(21)212018, (01)04056869299273(21)212041, (01)04056869299273(21)212038, (01)04056869299273(21)212047, (01)04056869299273(21)212019, (01)04056869299273(21)212037, (01)04056869299273(21)212032, (01)04056869299273(21)212049, (01)04056869299273(21)212017, (01)04056869299273(21)212064, (01)04056869299273(21)212096, (01)04056869299273(21)212065, (01)04056869299273(21)212016, (01)04056869299273(21)212023, (01)04056869299273(21)212056, (01)04056869299273(21)212054, (01)04056869299273(21)212063, (01)04056869299273(21)212051, (01)04056869299273(21)212055, (01)04056869299273(21)212021, (01)04056869299273(21)212060, (01)04056869299273(21)212011, (01)04056869299273(21)212066, (01)04056869299273(21)212074, (01)04056869299273(21)212090, (01)04056869299273(21)212034, (01)04056869299273(21)212029, (01)04056869299273(21)212097, (01)04056869299273(21)212099, (01)04056869299273(21)212039, (01)04056869299273(21)212059, (01)04056869299273(21)212030, (01)04056869299273(21)212035, (01)04056869299273(21)212036. Serial Numbers: 212087, 212098, 212088, 212068, 212081, 212018, 212041, 212038, 212047, 212019, 212037, 212032, 212049, 212017, 212064, 212096, 212065, 212016, 212023, 212056, 212054, 212063, 212051, 212055, 212021, 212060, 212011, 212066, 212074, 212090, 212034, 212029, 212097, 212099,…
    FDA recall number
    Z-2662-2025
    Quantity
    39 units

    The text above is the beginning of the record; 40 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.

    Lot or code
    1. Model Number: 10837643. UDI Number: (01)04056869006802(21)72030. Serial Number: 72030. Additional serial number added 11/14/2025: 72023. 2. Model Number: 10655588. UDI Number: (01)04056869006796(21)56069. Serial Number: 56069.
    FDA recall number
    Z-2663-2025
    Quantity
    3 units
  • MAGNETOM Verio. Model Number: 10276755.

    Lot or code
    Model Number: 10276755. UDI Numbers: N/A. Serial Numbers: 40649, 40585, 40781, 40243, 40408, 40628, 40848, 40870, 40783, 40473, 40240, 40497, 40564, 40425, 40778, 40389, 40340, 40523, 40582, 40168, 40291, 40283, 40201, 40850, 40331, 40126, 40744, 40821, 40433, 40502, 40179, 40246, 40273, 40766, 40753, 40828, 40737, 40352, 40313, 40382, 40670, 40387, 40631, 40669, 40131, 40541, 40466, 40292, 40375, 40161, 40745, 40268, 40284, 40325, 40799, 40190, 40877, 40104, 40195, 40407, 40640, 40797, 40254, 40142, 40665, 40622, 40264, 40420, 40633, 40150, 40514, 40727, 40710, 40328, 40156, 40800, 40692, 40165, 40441, 40864, 40482, 40472, 40859, 40220, 40858, 40779, 40346, 40295, 40233, 40629, 40681, 40464, 40568, 40200, 40663, 40334, 40277, 40276, 40879, 40222, 40361, 40553, 40378, 40714, 40715, 40252, 40250, 40423, 40204, 40486, 40251, 40488, 40109, 40855, 40429, 40232, 40157, 40440, 40259, 40545, 40556, 40687, 40140, 40237, 40399, 40312, 40235, 40697, 40684, 40398, 40871, 40700, 40368, 40656, 40606, 40163, 40278, 40562, 40671, 40526, 40658, 40644, 40376, 40686, 40713, 40775, 40863, 40326, 40409, 40116. Additional serial number added 11/14/2025: 40250. Additional serial number added 01/21/2026: 40459. Additional serial number added 2/25/2026: 40464. Additional serial number added 4/17/2026: 40711;
    FDA recall number
    Z-2664-2025
    Quantity
    154 units
  • MAGNETOM Verio Dot. Model Number: 10684333.

    Lot or code
    Model Number: 10684333. UDI Numbers: (01)04056869006673(21)60060, (01)04056869006673(21)160040, (01)04056869006673(21)160028, (01)04056869006673(21)160021, (01)04056869006673(21)60108, (01)04056869006673(21)60121, (01)04056869006673(21)160019, (01)04056869006673(21)60091, (01)04056869006673(21)160020, (01)04056869006673(21)160043, (01)04056869006673(21)60062, (01)04056869006673(21)60061, (01)04056869006673(21)160029, (01)04056869006673(21)160038, (01)04056869006673(21)60086, (01)04056869006673(21)60126, (01)04056869006673(21)60047, (01)04056869006673(21)160036, (01)04056869006673(21)160003, (01)04056869006673(21)160042, (01)04056869006673(21)160024, (01)04056869006673(21)60092, (01)04056869006673(21)160035, (01)04056869006673(21)60102, (01)04056869006673(21)160039, (01)04056869006673(21)60112, (01)04056869006673(21)160030, (01)04056869006673(21)60099, (01)04056869006673(21)160007, (01)04056869006673(21)160001, (01)04056869006673(21)160006, (01)04056869006673(21)160009, (01)04056869006673(21)160033, (01)04056869006673(21)60053, (01)04056869006673(21)60070, (01)04056869006673(21)60101, (01)04056869006673(21)60125, (01)04056869006673(21)60085, (01)04056869006673(21)60113, (01)04056869006673(21)160027, (01)04056869006673(21)160012, (01)04056869006673(21)160016, (01)04056869006673(21)160032, (01)04056869006673(21)160013, (01)04056869006673(21)160014, (01)04056869006673(21)160018, (01)04056869006673(21)160031, (01)04056869006673(21)160011, (01)04056869006673(21)160015,…
    FDA recall number
    Z-2665-2025
    Quantity
    59 units

    The text above is the beginning of the record; 771 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Verio Dot Upgrade. Model Number: 10684334.

    Lot or code
    Model Number: 10684334. UDI Numbers: (01)04056869006680(21)162012, (01)04056869006680(21)162055, (01)04056869006680(21)162039, (01)04056869006680(21)162002, (01)04056869006680(21)162014, (01)04056869006680(21)162069, (01)04056869006680(21)162083, (01)04056869006680(21)162075, (01)04056869006680(21)162028, (01)04056869006680(21)162060, (01)04056869006680(21)162035, (01)04056869006680(21)162038, (01)04056869006680(21)162016, (01)04056869006680(21)62040, (01)04056869006680(21)162036, (01)04056869006680(21)162079, (01)04056869006680(21)162054, (01)04056869006680(21)162088, (01)04056869006680(21)162071, (01)04056869006680(21)162078, (01)04056869006680(21)62032, (01)04056869006680(21)62037, (01)04056869006680(21)162077, (01)04056869006680(21)162042, (01)04056869006680(21)162025, (01)04056869006680(21)162082, (01)04056869006680(21)162085, (01)04056869006680(21)162026, (01)04056869006680(21)62027, (01)04056869006680(21)162040, (01)04056869006680(21)162032, (01)04056869006680(21)162072, (01)04056869006680(21)162037, (01)04056869006680(21)162043, (01)04056869006680(21)62014, (01)04056869006680(21)162047, (01)04056869006680(21)162023, (01)04056869006680(21)62019, (01)04056869006680(21)62050, (01)04056869006680(21)162089, (01)04056869006680(21)162061, (01)04056869006680(21)62018, (01)04056869006680(21)162004, (01)04056869006680(21)162080, (01)04056869006680(21)162013, (01)04056869006680(21)162007, (01)04056869006680(21)162049, (01)04056869006680(21)162031, (01)04056869006680(21)162009,…
    FDA recall number
    Z-2666-2025
    Quantity
    78 units

    The text above is the beginning of the record; 1,486 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.

    Lot or code
    Model Number: 11060815. UDI Numbers: (01)04056869039176(21)176636, (01)04056869039176(21)177211, (01)04056869039176(21)175728, (01)04056869039176(21)175685, (01)04056869039176(21)175851, (01)04056869039176(21)175768, (01)04056869039176(21)177298, (01)04056869039176(21)177295, (01)04056869039176(21)176756, (01)04056869039176(21)175754, (01)04056869039176(21)177349, (01)04056869039176(21)176757, (01)04056869039176(21)177232, (01)04056869039176(21)175935, (01)04056869039176(21)176967, (01)04056869039176(21)175741, (01)04056869039176(21)177052, (01)04056869039176(21)176060, (01)04056869039176(21)176371, (01)04056869039176(21)177012, (01)04056869039176(21)177146, (01)04056869039176(21)175770, (01)04056869039176(21)176375, (01)04056869039176(21)176122, (01)04056869039176(21)176308, (01)04056869039176(21)175669, (01)04056869039176(21)176212, (01)04056869039176(21)176755, (01)04056869039176(21)176781, (01)04056869039176(21)176302, (01)04056869039176(21)176036, (01)04056869039176(21)175766, (01)04056869039176(21)176825, (01)04056869039176(21)177151, (01)04056869039176(21)176087, (01)04056869039176(21)176771, (01)04056869039176(21)177299, (01)04056869039176(21)176451, (01)04056869039176(21)177352, (01)04056869039176(21)177354, (01)04056869039176(21)176779, (01)04056869039176(21)175870, (01)04056869039176(21)175838, (01)04056869039176(21)175692, (01)04056869039176(21)177359, (01)04056869039176(21)177119, (01)04056869039176(21)175977, (01)04056869039176(21)176037,…
    FDA recall number
    Z-2667-2025
    Quantity
    498 units

    The text above is the beginning of the record; 17,500 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MAGNETOM Vida Fit. Model Number: 11410481.

    Lot or code
    Model Number: 11410481. UDI Numbers: (01)04056869245188(21)202129, (01)04056869245188(21)202025, (01)04056869245188(21)202139, (01)04056869245188(21)202016, (01)04056869245188(21)202051, (01)04056869245188(21)202055, (01)04056869245188(21)202114, (01)04056869245188(21)202119, (01)04056869245188(21)202013, (01)04056869245188(21)202050, (01)04056869245188(21)202102, (01)04056869245188(21)202160, (01)04056869245188(21)202127, (01)04056869245188(21)202071, (01)04056869245188(21)202078, (01)04056869245188(21)202123, (01)04056869245188(21)202082, (01)04056869245188(21)202107, (01)04056869245188(21)202031, (01)04056869245188(21)202110, (01)04056869245188(21)202108, (01)04056869245188(21)202101, (01)04056869245188(21)202093, (01)04056869245188(21)202104, (01)04056869245188(21)202028, (01)04056869245188(21)202026, (01)04056869245188(21)202076, (01)04056869245188(21)202142, (01)04056869245188(21)202125, (01)04056869245188(21)202155, (01)04056869245188(21)202084, (01)04056869245188(21)202151, (01)04056869245188(21)202068, (01)04056869245188(21)202135, (01)04056869245188(21)202121, (01)04056869245188(21)202006, (01)04056869245188(21)202010, (01)04056869245188(21)202011, (01)04056869245188(21)202146, (01)04056869245188(21)202145. Serial Numbers: 202129, 202025, 202139, 202016, 202051, 202055, 202114, 202119, 202013, 202050, 202102, 202160, 202127, 202071, 202078, 202123, 202082, 202107, 202031, 202110, 202108, 202101, 202093, 202104, 202028, 202026, 202076, 202142, 202125, 202155, 202084,…
    FDA recall number
    Z-2668-2025
    Quantity
    41 units

    The text above is the beginning of the record; 125 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.

Source

Issued by
U.S. Food and Drug Administration
Firm
Siemens Medical Solutions USA, Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Siemens Medical Solutions USA, Inc, Month: October 2025, Week 2025-W41. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97530 on October 8, 2025.

Other notices of Siemens Medical Solutions USA, Inc