What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
ARTIS Pheno VE30A and VE40A, Model 10849000
- Lot or code
- UDI 04056869046877
- FDA recall number
- Z-1814-2026
- Quantity
- 73
Distribution (FDA): U.S. and O.U.S.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Siemens Medical Solutions USA, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.