What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
- Lot or code
- ARTIS pheno UDI: 4056869046877. ARTIS icono biplane UDI: 4056869063317. ARTIS icono ceiling UDI: 4056869295923. ARTIS icono floor UDI: 4056869149325.
- FDA recall number
- Z-1653-2026
- Quantity
- U.S. 758, OUS 2077
Distribution (FDA): Worldwide - US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Siemens Medical Solutions USA, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.