What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MAGNETOM Cima.X Upgrade. Model Number: 11689304.
- Lot or code
- Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.
- FDA recall number
- Z-0431-2026
- Quantity
- 4 system (1 US; 3 OUS)
Distribution (FDA): US distribution to California. International distribution to Australia, Germany, Great Britian.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Siemens Medical Solutions USA, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.