Medtronic MiniMed, Inc. recalls
15 notices naming Medtronic MiniMed, Inc. as the recalling firm, since March 12, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Medtronic MiniMed, Inc.: Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600
Hazard (FDA): A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating…
FDAMedical devices
Medtronic MiniMed, Inc.: CareLink Clinic
Hazard (FDA): CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily…
FDAMedical devices
Medtronic MiniMed, Inc.: Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers…
Hazard (FDA): Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed 640G Insulin Pump and 2 more
Hazard (FDA): Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed Go App
Hazard (FDA): For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump
Hazard (FDA): A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed 670G Insulin Pump and 13 more
Hazard (FDA): All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
Hazard (FDA): Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or…
FDAMedical devices
Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60…
Hazard (FDA): A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.
FDAMedical devices
Medtronic MiniMed, Inc.: InPen App
Hazard (FDA): When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on…
FDAMedical devices
Medtronic MiniMed, Inc.: CareLink Clinic
Hazard (FDA): Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview…
FDAMedical devices
Sterility
Medtronic MiniMed, Inc.: Medtronic
Hazard (FDA): The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
FDAMedical devices
Medtronic MiniMed, Inc.: InPen App
Hazard (FDA): Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose…
FDAMedical devices
Medtronic MiniMed, Inc.: InPen smart insulin pen
Hazard (FDA): Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove…
FDAMedical devices
Medtronic MiniMed, Inc.: Paradigm REF: MMT-712 and MMT-715 and 11 more
Hazard (FDA): All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane…
FDAMedical devices