Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96200: Medtronic MiniMed, Inc.: Paradigm REF: MMT-712 and MMT-715 and 11 more

FDA recall 96200 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing of the 780G NGP infusion pump under dynamic pressure conditions, the firm observed that when air pressure decreased during take-off simulation, on average 0.569U more insulin was delivered in 15 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate; when air pressure increased during landing simulation, on average 0.537 U less insulin was delivered in 25 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate. The firm also observed that the mechanisms causing the variations in delivery volumes are independent of the delivery rate. All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

32 more codes are in the record below.

  • Paradigm REF: MMT-712 and MMT-715

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1255-2025
    Quantity
    20,760 units
  • Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1256-2025
    Quantity
    46,437 units
  • Paradigm insulin pump, REF: MMT-754

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1257-2025
    Quantity
    41,404 units
  • MiniMed insulin pump, REF: 620G (MMT-1750)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1258-2025
    Quantity
    228 units
  • MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1259-2025
    Quantity
    70587 units
  • MiniMed insulin pump, REF: 640G (MMT-1711, MMT-1712, MMT-1751, MMT-1752)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1260-2025
    Quantity
    168,240
  • MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1261-2025
    Quantity
    100,074 units
  • MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1262-2025
    Quantity
    4,959 unit
  • MiniMed insulin pump, REF: 720G (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1263-2025
    Quantity
    8,677
  • MiniMed insulin pump, REF: 740G (MMT-1811, MMT-1812, MMT-1861, MMT-1862)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1264-2025
    Quantity
    15,439 units
  • MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1265-2025
    Quantity
    106,289 units
  • MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)

    Lot or code
    All Lot Numbers; All UDI numbers
    FDA recall number
    Z-1266-2025
    Quantity
    329,025 units

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Algeria, Andorra, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State Of Bosnia, Herzegovina , Botswana, Brazil, Brunei, Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Cayman, Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic Of Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malawi, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Republic Of Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russia, Saint Pierre And Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Tanzania, United Republic Of Thailand, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic Of Vietnam, Yemen, Zimbabwe.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medtronic MiniMed, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medtronic MiniMed, Inc., Month: March 2025, Week 2025-W11. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing of…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96200 on March 12, 2025.

Other notices of Medtronic MiniMed, Inc.