What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
280 more codes are in the record below.
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Lot or code
- Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H
- FDA recall number
- Z-2980-2026
- Quantity
- 16
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Lot or code
- Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H
- FDA recall number
- Z-2981-2026
- Quantity
- 7
MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW
- Lot or code
- Model/UDI-DI Serial Numbers: MMT-1885CLW/00763000442187, 00763000974374/NG3150737H, NG3155194H, NG3151561H, NG3154693H, NG3151716H, NG3153547H, NG3153637H, NG3155407H, NG3152666H, NG3155788H, NG3081975H, NG2901117H, NG3080842H, NG3081463H, NG3153416H, NG3081894H, NG3153781H, NG3155944H, NG3081982H, NG3081525H, NG3155909H, NG3090722H, NG3154897H, NG3088436H, NG3088708H, NG3090073H, NG3155908H, NG3150720H, NG3092957H, NG3128715H, NG3128784H, NG3152774H, NG3153929H, NG3514566H, NG3150780H, NG3151290H, NG3151508H, NG3507161H, NG3508693H, NG3081976H, NG3151460H, NG3152051H, NG3152085H, NG3152344H, NG3153213H, NG3152772H, NG3155586H, NG3150786H, NG3153214H, NG3153385H, NG3151840H, NG3082019H, NG3153405H, NG3153788H, NG3153908H, NG3154527H, NG3151598H, NG2854971H, NG3155196H, NG3155202H, NG3155341H, NG3091491H, NG3155557H, NG3081498H NG3155688H, NG3155020H, NG3508565H, NG3155419H, NG3507536H, NG3091563H, NG3082036H, NG3154689H, NG3154995H, NG3508504H, NG3514345H, NG3509850H, NG3527505H, NG3514547H, NG3527469H, NG3504333H, NG3504598H, NG3505071H, NG3505220H, NG3505273H, NG3514617H, NG3506727H, NG3085044H, NG3507542H, NG3507547H, NG3514448H, NG3508303H, NG3508350H, NG3508371H, NG3514612H, NG3508511H, NG3508617H, NG3508698H, NG3508707H, NG3508757H, NG3508799H, NG3508903H, NG3509138H, NG3509295H, NG3509299H, NG3509452H, NG3509685H, NG3509689H, NG3509836H, NG3509914H, NG3509925H, NG3514375H, NG3514441H, NG3505246H, NG3505166H, NG3505283H, NG3151570H, NG3154948H, NG3514667H, NG3527165H,…
- FDA recall number
- Z-2982-2026
- Quantity
- 257
The text above is the beginning of the record; 1,697 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medtronic MiniMed, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.