Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98186: Medtronic MiniMed, Inc.: MiniMed 670G Insulin Pump and 13 more

FDA recall 98186 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

388 more codes are in the record below.

  • MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 670G Insulin Pump: 643169946057, 763000096984, 763000090203, 763000165468, 763000190460, 763000190453, 763000216481, 763000187422, 763000187415, 763000165413, 763000165475, 763000367657, 763000289065, 763000264710, 763000283476, 763000283544, 763000283520, 763000283445, 763000393786, 763000289058, 763000374532, 763000283537, 763000365899, 763000371937, 763000283513, 763000165451, 763000165420, 763000165444, 763000165437, 763000165499, 763000165482, 763000165536, 763000165512, 763000165550, 763000165529, 763000235772, 763000165505, 763000235765, 763000165543, 763000257309, 763000283568, 763000283599, 763000289041, 763000289089, 763000289195, 763000289072, 763000289119, 763000289188, 763000289133, 763000289126, 763000289171, 763000289102, 763000404543, 763000289096, 763000289164, 763000404536, 763000324797, 763000264727, 763000324780, 763000283551, 763000283582, 763000374518, 763000374525, 763000374556, 763000374624, 763000374594, 763000374600, 763000399849, 763000399856, 763000408916, 763000408947, 763000371944, 763000365905, 763000374587, 763000408930, 763000283575, 763000157463, 763000203313, 763000252076, 763000283506, 763000411787, 643169946057, 763000096984, 763000190453, 763000283445, 763000395858, 763000278427, 763000090203, 763000283513, 763000365882, 763000283506, 76300028351303, 763000393786, 76300028351301, 763000166526, 643169939202, 763000190460, 763000283520, 763000072537, 763000256821, 643169939219,…
    FDA recall number
    Z-1739-2026
    Quantity
    84,248

    The text above is the beginning of the record; 609 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 630G Insulin Pump: 763000253066, 643169752726, 763000101527, 763000190446, 763000317591, 763000157043, 643169752726, 763000101527, 763000190446, 763000317591, 643169873841, 643169878648, 763000179632, 76300031759101, 76300031759103, 76300031762103, 763000367091, 763000317621, 763000157043, 76300031762101: MiniMed 630G System(M020092C004 1); MINIMED 630G and MINIMED 670G INSULIN PUMP SYSTEM(M042888C002 1); MiniMed 630G Insulin pump MODEL MMT-1714(M993428A262 1); MiniMed 630G System User Guide(M020092C001 3, M020092C003 1, M020092C005 1); MINIMED 630G SYSTEM USER GUIDE mg/dL(M979512A011 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE(MP6026120-262 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE(M030278C001 3, M030278C002 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE mmol/L (MP6026120-024 1); MiniMed Veo with Enlite Threshold Suspend Performance(M10577668A262 1).
    FDA recall number
    Z-1740-2026
    Quantity
    60,880
  • MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 780G Insulin Pump: 763000411541, 763000441449, 763000523305, 763000521530, 763000521516, 763000533861, 763000544317, 763000734459, 763000758530, 763000758547, 763000578688, 763000734435, 763000504762, 763000734466, 763000924102, 763000924089, 763000454296, 763000854829, 763000411411, 763000411503, 763000384272, 763000411466, 763000411428, 763000411572, 763000411589, 763000411558, 763000384265, 763000411473, 763000411374, 763000411435, 763000411404, 763000384289, 763000411459, 763000411510, 763000411626, 763000411565, 763000411527, 763000411619, 763000411596, 763000411497, 763000411398, 763000411534, 763000411381, 763000384258, 763000411480, 763000411602, 763000411640, 763000411442, 763000441432, 763000442194, 763000521523, 763000442187, 763000521509, 763000521547, 763000578671, 763000596194, 763000639228, 763000639235, 763000639266, 763000974381, 763000974374, 763000921750, 763000921767, 763000921590, 763000921637, 763000933593, 763000891886, 763000921729, 763000921866, 763000921781, 763000921811, 763000921576, 763000921583, 763000921620, 763000921859, 763000891909, 763000921743, 763000944162, 763000921712, 763000921668, 763000933579, 763000933555, 763000933586, 763000933609, 763000921613, 763000933715, 763000933548, 763000944209, 763000891893, 763000933623, 763000933722, 763000933814, 763000933708, 763000933777, 763000933685, 763000933531, 763000933678, 763000933821, 763000933746, 763000933562, 763000944193,…
    FDA recall number
    Z-1741-2026
    Quantity
    453,144

    The text above is the beginning of the record; 5,163 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 770G Insulin Pump: 763000456634, 763000414344, 763000413750, 763000545604, 763000586188, 763000642242, 763000926113, 763000482602, 763000416553, 763000439903, 763000416546, 763000439910, 763000629892, 763000629908, 00763000595906, 763000412685, 763000418151, 763000416546, 763000439903, 763000545499, 763000595906, 763000545505, 763000629892, 763000642235, 763000414344, 763000413750, 763000439866, 763000413750, 763000586195, 763000545604, 763000578640, 76300058618802, 76300058619501, 76300057864002, 76300054560402, 76300058619502, 763000926113: For MiniMed 770G System Users Ages 2-13(M010831C001 1); MiniMed 770G SYSTEM USER GUIDE(M984970A002 1, M978700A002 1, M030281C001 3, M030281C002 1, M811470A261 2, M984970A004 1, M020090C001 3, M020090C002 1, M020090C003 1, M998590A012 2, M998590A003 1, M998590A004 1); MiniMed 770G SYSTEM USER GUIDE mmol/L(M000039C021 1, M811470A021 2).
    FDA recall number
    Z-1742-2026
    Quantity
    95,704
  • MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 640G Insulin Pump: 643169999763, 763000002763, 763000002671, 643169999770, 763000002886, 643169729353, 763000002657, 763000002688, 763000002664, 763000002749, 763000002817, 763000002787, 763000002800, 763000002756, 763000002848, 763000002794, 763000002695, 763000002725, 763000002701, 763000002633, 763000002862, 643169729360, 763000002879, 763000002824, 763000002732, 763000002640, 763000002770, 763000002855, 763000002718, 763000002831, 763000065195, 763000065201, 763000083441, 763000015510, 763000015541, 763000015480, 763000015589, 763000015497, 763000068998, 763000015381, 763000015435, 763000015336, 763000015367, 763000069001, 763000015374, 763000015398, 763000015350, 763000192228, 763000015572, 763000015343, 763000015411, 763000015428, 763000015404, 763000015442, 763000015459, 763000015534, 763000015503, 763000194383, 763000194222, 763000194444, 763000194390, 763000194239, 763000198596, 763000194406, 763000194369, 763000015596, 763000194277, 763000194666, 763000194352, 763000198602, 763000192211, 763000194451, 763000015565, 763000194246, 763000194338, 763000194307, 763000194475, 763000194314, 763000155346, 763000194468, 763000194437, 763000194376, 763000194345, 763000194253, 763000194260, 763000194413, 763000194420, 763000192136, 763000192143, 763000165246, 763000165307, 763000165130, 643169629813, 763000194321, 763000194291, 763000165147, 763000165277, 763000165222, 763000165116, 763000165123, 763000165185,…
    FDA recall number
    Z-1743-2026
    Quantity
    116,645

    The text above is the beginning of the record; 1,792 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 700G Insulin pump: 763000552541, 763000541323, 763000854737, 763000854720: MINIMEDTM 700 INSULIN PUMP USER GUIDE mmol/L(M984972A021 1); MINIMED 700 INSULIN PUMP USER GUIDE mg/dL(M015850C001 1).
    FDA recall number
    Z-1744-2026
    Quantity
    13,811
  • MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 720G Insulin Pump: 763000854768, 763000505196, 763000505219, 763000505318, 763000505257, 763000505295, 763000505233, 763000505370, 763000505325, 763000505332, 763000505394, 763000505356, 763000505387, 763000505349, 763000505400, 763000505226, 763000505271, 763000638016, 763000531171, 763000854751: MINIMED 720G SYSTEM USER GUIDE(M000187C022 1, M000186C002 1, M000186C014 1, M000187C016 1, M000186C021 1, M000186C018 1, M000186C017 1, M000187C010 1, M000187C011 1, M000186C031 1, M000187C013 1, M000187C008 1, M000187C015 1, M000187C005 1, M000187C033 1, M000186C028 1, M000187C023 1, M000187C020 1, M000187C018 1, M000186C018 13, M000186C004 1, M000186C030 1, M000186C009 1, M000187C024 1, M000186C033 1, M000186C003 1, M000187C012 1).
    FDA recall number
    Z-1745-2026
    Quantity
    37,458
  • MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 740G Insulin Pump: 763000504908, 763000504915, 763000505073, 763000484637, 763000505042, 763000504861, 763000629984, 763000504946, 763000504878, 763000559557, 763000504960, 763000505080, 763000505066, 763000629991, 763000505004, 763000504922, 763000504892, 763000505059, 763000504977, 763000505028, 763000505172, 763000505127, 763000505134, 763000505158, 763000505141, 763000505103, 763000505110, 763000504991, 763000505165, 763000504939, 763000504953, 763000504984, 763000505189, 763000851224, 763000851217, 763000769451, 763000769468: MINIMED 740G SYSTEM USER GUIDE mg/dL(M000149C002 1, M000149C004 1, M000149C005 1, M000150C005 1, M000150C002 1, M000150C021 2, M000149C016 1, M000149C021 1, M000150C004 1, M000150C025 1, M000150C030 1, M000149C014 1, M000149C006 1, M000150C003 1, M000150C015 1, M000150C028 1, M000150C018 1, M000150C022 1, M000150C007 1, M000149C017 1, M000149C008 1, M000150C008 1, M000150C032 1, M000150C029 1, M000150C012 1, M000150C016 1, M000150C010 1, M000149C011 1, M046726C001 3, M046726C002 3, M000149C020 1, M000149C025 1, M000150C024 1, M000150C023 1, M000150C013 1).
    FDA recall number
    Z-1746-2026
    Quantity
    24,946
  • MiniMed 620G Insulin Pump (MMT-1750)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 620G Insulin Pump: 643169559745, 643169721319, 763000375461, 763000253189: Instr, Startup Guide, 620G 4.10, EN(6025951-022 1); INSTR., NGP 620G, BOLUS WIZARD, EN(6026006-023 1); Instr., 620G, Sftwr. 4.10, EN.(6026006-024 1); INSTRUCTIONS MMT-746EN PUMP 1510/1710 EN(7005933-023 1); DOC PKG MMT-746EN NGP 1710 EN(7005933-024 1); Startup Guide Assembly MMT-755EN(620G(M982438A021 1).
    FDA recall number
    Z-1747-2026
    Quantity
    799
  • Paradigm Insulin Pump (MMT-712, MMT-715)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Insulin Pump: 643169999770, 643169729360, 763000065201, 763000069001, 763000192228, 763000198602, 763000155346, 763000192143, 763000258344, 763000317140, 763000404567, 763000317188, 763000317195, 763000367046, 763000317157, 00613994470492, 00643169496262, 643169577664, 643169577688, 643169577695, 643169739512, 00613994332929, 00613994469694, 00613994470904, 00763000084271, 00763000161019, 643169577701, 643169739512, 643169662612, 00643169533776, 00763000084370, 643169656840, 763000090197, 763000190439, 763000316631, 00613994473820, 00643169478503, 00613994151766, 00643169710368, 00763000161057, 00613994326478, 00613994326508, 00613994395771, 00613994395719, 00613994395696, 00613994527851, 00643169729421, 00643169198920, 00643169572447, 00643169198890, 00613994527868, 00613994172129, 00613994172419, 00613994171627, 00613994171344, 00613994172211, 00613994172525, 00613994172334, 00613994172044, 00613994172761, 00613994171917, 00613994326621, 00613994326584, 00643169848825, 00643169848832, 00643169848801, 00613994395818, 00613994395764, 00613994395740, 00613994394132, 00763000069209, 00763000069179, 00613994527905, 00763000069162, 00613994527769, 00613994405876, 00763000069186, 00763000069148, 00613994527783, 00613994527806, 00613994527899, 00613994527790, 00613994527882, 00613994527837, 00643169198968, 00643169198883, 00643169198944, 00643169198906, 00763000084301, 00613994527745, 00643169198975, 00643169035515,…
    FDA recall number
    Z-1748-2026
    Quantity
    73,656

    The text above is the beginning of the record; 913 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Insulin Pump: 00613994409423, 00613994622600, 613994622747, 613994407733, 00613994407740, 00613994136374, 00613994271600, 00613994106476, 00613994373120, 00613994409478, 00613994312426, 00613994151780, 00885074419831, 00613994622747, 00613994951502, 00613994622648, 00643169026100, 00613994622624, 00613994622556, 00613994402011, 00613994409553, 00613994368485, 00613994368508, 00613994368522, 00613994108739, 00613994368607, 00613994368492, 00613994368515, 00613994123084, 00613994124814, 00613994402523, 00613994402530, 00613994402516, 00613994401502, 00613994402509, 00613994405913, 00613994123206, 00643169688421, 00643169199699, 00643169199675, 00613994409461, 00613994124777, 00643169572492, 00613994267177, 00643169848054, 00613994266897, 00613994310804, 00613994310798, 00613994310774, 00613994310903, 00613994401564, 00613994401663, 00613994402554, 00613994402462, 00613994402592, 00613994402608, 00613994405944, 00613994402455, 00613994402547, 00613994402479, 00763000069292, 00613994402578, 00613994409447, 00643169729438, 00763000069254, 00763000084325, 00613994402486, 00613994405920, 00763000069261, 00613994948595, 00613994929778, 00613994224569, 00613994929808, 00613994930361, 00613994929761, 00643169199705, 00643169199668, 00643169496248, 00643169848016, 00643169848023, 00763000161095,00613994123329, 00613994125026, 00613994247728, 00613994124692, 00613994124739, 00613994266903, 00613994267153,…
    FDA recall number
    Z-1749-2026
    Quantity
    29,074

    The text above is the beginning of the record; 1,623 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Revel Insulin Pump: 00613994951502, 00763000161262, 00613994743930, 00643169077737, 00613994904485, 00643169202399, 00643169151369, 00643169478602, 00643169513914, 00643169933170, 00763000161132, 00643169464520: Artwork, CRB D Process, Paradigm REAL-Time Revel Insulin Pump UG (MP6025306-011A 1); Artwork, CRB D Process, Instr, X23 revel, EN (MP6025306-019 1)
    FDA recall number
    Z-1750-2026
    Quantity
    26,647
  • MiniMed 530G Insulin Pump (MMT-551, MMT-751)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 530G: 00643169507784, 00643169736061, 00643169736092, 00643169736085, 00643169933224, 00763000161170: Artwork, CRB A Process, Instr, 530G, One Press, Launch, US(MP6026081-012DOC); Artwork, CRB A Process, Instr, 530G, FDA comments, add TST(MP6025813-013DOC); Artwork, CRB A Process, Instr, 530G, Remove Velcro, Icons, US(MP6025813-014DOC);
    FDA recall number
    Z-1751-2026
    Quantity
    13,429
  • Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)

    Lot or code
    Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Veo: 00613994717801, 00613994713834, 00613994713841, 00613994870414, 00613994870377, 00613994870407, 00643169378636, 00613994924957, 00613994717672, 00613994715531, 00613994715555, 00613994715562, 00613994409423, 00613994951502, 00643169026131, 00613994622600, 00613994407740, 00643169077737, 00643169464537, 00643169305038, 00643169500259, 00643169503229, 00643169528789, 00643169533776, 00613994151766, 00643169711440, 00763000084370, 00763000161224, 00763000161262, 00643169464520, 00613994963376, 00643169026100, 00613994924247, 00643169299757, 00643169400931, 00643169400986, 00643169535749, 00643169535794, 00643169519190, 00643169579507, 00643169519091, 00763000084349, 00763000084356, 00763000161248, 00763000048853, 00763000161231, 00643169729445, 00643169846036, 00643169579453, 00613994714183, 00613994716156, 00613994657671, 00643169515741, 00643169579514, 00643169579538, 00643169579477, 00643169579569, 00643169579392, 00643169579583, 00643169579552, 00643169579576, 00643169846012, 00643169845909, 00643169845947, 00643169845992, 00643169846043, 00763000067663, 00643169846005, 00763000001247, 00643169578470, 00763000059729, 00763000048860,00763000048914, 00763000048839, 00763000048884, 00763000048822, 00763000048808, 00763000048938, 00763000048815, 00763000048921, 00763000048891, 00763000048785, 00763000048945, 00763000048877, 00763000048846, 00613994924438, 00613994924223, 00613994924391, 00613994924339,…
    FDA recall number
    Z-1752-2026
    Quantity
    55,724

    The text above is the beginning of the record; 2,908 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medtronic MiniMed, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medtronic MiniMed, Inc., Month: April 2026, Week 2026-W15. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98186 on April 8, 2026.

Other notices of Medtronic MiniMed, Inc.