What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
9 more codes are in the record below.
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
- Lot or code
- UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
- FDA recall number
- Z-2698-2026
- Quantity
- 778
Distribution (FDA): International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medtronic MiniMed, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.