CareFusion 303, Inc. recalls
29 notices naming CareFusion 303, Inc. as the recalling firm, since December 18, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Burn
CareFusion 303, Inc.: BD Pyxis MedStation 4000 REF: 303 Medication cabinet and 7 more
Hazard (FDA): Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
FDAMedical devices
CareFusion 303, Inc.: BD Alaris" System with Guardrails" Suite MX with Point of Care Unit
Hazard (FDA): Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
FDAMedical devices
CareFusion 303, Inc.: BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System…
Hazard (FDA): Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis ES Enterprise Server Catalog Numbers 1115-00
Hazard (FDA): Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis MedStation ES
Hazard (FDA): Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to…
FDAMedical devices
CareFusion 303, Inc.: BD Alaris infusion Pump Module 8100
Hazard (FDA): If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred…
FDAMedical devices
Fire
CareFusion 303, Inc.: Various models of BD Pyxis Medbank: Reference numbers: 1137-00 1145-00 1146-00 1147-00… and 4 more
Hazard (FDA): Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
FDAMedical devices
CareFusion 303, Inc.: BD Alaris Pump Infusion Sets for the following reference numbers: C24101E
Hazard (FDA): Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main /…
Hazard (FDA): Antivirus software was not consistently installed on impacted devices during the implementation process.
FDAMedical devices
CareFusion 303, Inc.: Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137
Hazard (FDA): Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
FDAMedical devices
CareFusion 303, Inc.: BD Alaris Pump Module Model and 1 more
Hazard (FDA): Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion…
FDAMedical devices
CareFusion 303, Inc.: Product Name [REF]: Supply Server
Hazard (FDA): During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the…
FDAMedical devices
CareFusion 303, Inc.: BD PYXIS MEDBANK
Hazard (FDA): Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock…
FDAMedical devices
CareFusion 303, Inc.: MEDFLEX
Hazard (FDA): If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower
Hazard (FDA): Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR…
Hazard (FDA): Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis: Anesthesia Station ES
Hazard (FDA): Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing…
FDAMedical devices
CareFusion 303, Inc.: BD Alaris Systems Manager and 1 more
Hazard (FDA): Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis: MedFlex
Hazard (FDA): Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result…
FDAMedical devices
CareFusion 303, Inc.: 138913-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 BD PYXIS MEDBANK MINI CR-2HH-1FH-P… and 3 more
Hazard (FDA): Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
FDAMedical devices
CareFusion 303, Inc.: BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a…
Hazard (FDA): Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the…
FDAMedical devices
CareFusion 303, Inc.: BD Pyxis MedFlex 2.0 REF 1139-00 Bio-ID Spare Parts REF 353200-01 The BD Pyxis" MedFlex…
Hazard (FDA): Due to fingerprint scanner failing resulting in the scanner heating up.
FDAMedical devices
CareFusion 303, Inc.: (1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS)…
Hazard (FDA): Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.
FDAMedical devices
CareFusion 303, Inc.: (1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog…
Hazard (FDA): Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues
FDAMedical devices
- FDACareFusion 303, Inc.: BD Pyxis MedStation ES and 3 more
- FDACareFusion 303, Inc.: BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314…
- FDACareFusion 303, Inc.: BD Pyxis MedStation ES and 5 more
- FDACareFusion 303, Inc.: BD Pyxis MedStation ES
- FDACareFusion 303, Inc.: BD Pyxis MedStation ES and 2 more