Why it was recalled

During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

3 more codes are in the record below.

  • Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main 10885403512629 [303], BD Pyxis MedStation ES Tower 10885403512674 [352], BD Pyxis Anesthesia Station ES 10885403477836 [327], BD Pyxis MedStation ES (Main) 10885403512667 [323], MD BD Pyxis SupplyStation V9 [317], Security Module [139698-01], BD Knowledge Portal for Medication Technologies [136607-01], STOCKSTN V10.X WRD/WL SCANNERS [132-53-02], PRS STOCKSTN V10.X WRD/WL SCANNERS [132-53-01], STOCKSTN V10.X WIRED SCANNER [132-52-02], PRS STOCKSTN V10.X WIRED SCANNER [132-52-01], BD Pyxis CII Safe ES Desktop PC [1156-00], BD Pyxis SupplyStation Panel PC V11 [1132-00], BD Pyxis SupplyRoller V11 [1127-00], Dell 640 Server [1128-00], BD Pyxis SupplyStation System V11 [1123-00], BD Pyxis CII Safe ES 10885403512605 [1116-00], BD Pyxis Enterprise Server 10885403518348, 10885403520341 [1115-00], BD Pyxis SupplyRoller V10 [1101-00], BD Pyxis CIISafe V9.X SGL IM BIO 10885403512520 [111-221], BD Pyxis CIISafe V7.X DBL INTG MAIN BIO, SOLID DO [111-194], BD Pyxis CIISafe V8.X Desktop PC, Bio [107-245-01], BD Pyxis SupplyStation V10 [347], BD Pyxis CII Safe V9.X DESKTOP PC [107-252-01], BD Pyxis CII Safe V7.X DBL INTG MAIN BIO CLEAR DO [111-191], BD Pyxis CII Safe V7.XSGL INTG MAIN BIO SOLID DO [111-198], BD Pyxis CII SafeV8.XDBL INTG MAIN BIO SOLID DO [111-202-01], BD Pyxis CII SafeV8.XSGL INTG MAIN BIO SOLID DO [111-206-01], BD Pyxis CII Safe V9.X DBL IM BIO…

    Lot or code
    Each customer was provided a unique list of their affected devices that included computer names and atlas keys. Distributed: April 14, 2014 - March 21, 2025 REF or Name [UDI-DI]: Supply Server [N/A], CCE Enterprise SW Site License (2-5) [N/A], 806 [N/A], 303 [10885403512629], 352 [10885403512674], 327 [10885403477836], 323 [10885403512667], 317 [N/A], 139698-01 [N/A], 136607-01 [N/A], 132-53-02 [N/A], 132-53-01 [N/A], 132-52-02 [N/A], 132-52-01[N/A] , 1156-00 [N/A], 1132-00 [N/A], 1127-0 [N/A], 1128-00 [N/A], 1123-00 [N/A], 1116-00 [N/A], 1115-00 [10885403518348, 10885403520341], 1101-00 [N/A], 111-221 [10885403512520], 111-194 [N/A], 107-245-01 [N/A], 347[N/A], 107-252-01 [N/A], 111-191 [N/A], 111-198 [N/A], 111-202-01 [N/A], 111-206-01 [N/A], 111-215 [10885403512520], 111-216 [10885403512520], 111-219 [10885403512520], 138566-01 [N/A]
    FDA recall number
    Z-2017-2025
    Quantity
    6,332

    The text above is the beginning of the record; 185 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of PR, DE, GU, MO, IA, IL, CA, MD, MA, VA, PA, NY, WA, NJ, AZ, CO, OH, FL, WI, TX, CT, LA, MI, WY, MN, AR, KY, OR, NM, TN, NC, SC, IN, VT, KS, AL, GA, WV, NV, ID, DC, MS, NE, HI, OK, UT, ND, MT, NH, SD, ME, RI, AK, MP and the countries of SA, CA, NZ , CH, FR, IT, BE, AU, BH, QA, AE, GB, EG, AR, JO, DE, BM, TH, LB, BR, IE, FI, MX, TW, ID, SG, BS, IN, MA, GF, ES, HK.

Source

Issued by
U.S. Food and Drug Administration
Firm
CareFusion 303, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: CareFusion 303, Inc., Month: July 2025, Week 2025-W27. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96585 on July 2, 2025.

Other notices of CareFusion 303, Inc.