Skip to content
FDAMedical devicesClass IOngoing

FDA recall 97519: CareFusion 303, Inc.: BD Alaris Pump Infusion Sets for the following reference numbers: C24101E

FDA recall 97519 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

10 more codes are in the record below.

  • BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way Stopcock

    Lot or code
    All Lots/ UDI: C24101E - 10885403235528, 10015414 - 10885403233999, 2447-0007 - 10885403235078, 24201-0007 - 10885403476341, 10012645 - 10885403233753, 2426-0500 - 7613203020992, 11522558 - 10885403232305, 2410-0500 - 7613203012492, 2232-0007 - 10885403463969, 11613191 - 10885403276026, 10012144 - 7613203021159, 10013034 - 10885403232312, 10013037 - 10885403233784, 11582773 - 10885403239588
    FDA recall number
    Z-0035-2026

Distribution (FDA): US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa

Source

Issued by
U.S. Food and Drug Administration
Firm
CareFusion 303, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: CareFusion 303, Inc., Month: October 2025, Week 2025-W42. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97519 on October 15, 2025.

Other notices of CareFusion 303, Inc.