What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
(1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis CII Safe ES, Catalog No. 1116-00
- Lot or code
- (1) Pyxis MedStation ES, Catalog No. 323, UDI-DI 10885403512667; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327, UDI-DI 10885403477836; (3) Pyxis MedStation ES Tower, Catalog No. 352, UDI-DI 10885403512674; (4) BD Pyxis CII Safe ES, Catalog No. 1116-00, UDI-DI 10885403512605; All Serial Numbers; Software Versions v1.10 and prior Affected
- FDA recall number
- Z-1119-2025
- Quantity
- 137,741 total
Distribution (FDA): Domestic distribution nationwide. International distribution to North America, Latin America, EMEA, and Greater Asia.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- CareFusion 303, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.