Why it was recalled

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • BD Pyxis MedStation ES, REF: 323

    Lot or code
    All software versions. REF/UDI-DI: 323/10885403512667/ 12441103 12440974 12440975 12441333 12441432 12447684 12455493 12455498 12455492 12455793 12458864 12463612 12470842 12476547 12482532 12483393 12486737 12499043 12499838 12499616 12499617 12499618 12505982 12508303 12514216 12514218 12514217 12514220 12515358 12522942 12527930 12528710 12528709 12530090 12531418 12537364 12537363 12537076 12540196 12540194 12540267 12540268 12540269 12561139 12568697 12569170 12570749 12570750 12570745 12570746 12570748 12570753 12585184 12585183 12586285 12589659 12589685 12589658 12589618 12589657 12593259 12594318 12594322 12596587 12596404 12597477 12597478 12597479 12597474 12597489 12597491 12597492 12597493 12597475 12598838 12599570 12599571 12599063 12599064 12599065 12599067 12599069 12599066 12599070 12599061 12599062 12599135 12599136 12599138 12599140 12599137 12599141 12599139 12599971 12600268 12599970 12600435 12600467 12600468 12600469 12600690 12600687 12600689 12600691 12602140 12602358 12602360 12602357 12602359 12602372 12602361 12605985 12606524 12608631 12611808 12611816 12611926 12612669 12612567 12615383 12616239 12624466 12626212 12629311 12629309 12629602 12629604 12629603 12629627 12630890 12632195 12633312 12638031 12637962 12637963 12637961 12638017 12638018 12639538 12650679 12650091 12652625 12653147 12653602 12655694 12656692 12656693 12656812 12658185 12659515 12662647 12662921 12666199 12666200 12676117 12678689 12681469 12681569 12683904 12685086…
    FDA recall number
    Z-0670-2025
    Quantity
    98,526

    The text above is the beginning of the record; 885,429 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324

    Lot or code
    All software versions. REF/UDI-DI: 324/10885403512681/ 12440840 12440841 12448208 12452252 12452264 12455476 12462734 12462735 12463676 12468609 12468612 12468607 12468608 12468610 12468671 12468672 12468674 12468670 12468673 12468611 12474015 12474022 12473810 12473813 12473815 12473816 12473817 12473811 12473921 12474012 12474071 12474014 12473887 12473922 12474013 12473807 12473812 12473808 12473809 12476548 12495456 12499285 12502858 12502857 12503069 12508589 12508584 12508578 12508579 12508587 12508580 12510555 12510559 12510393 12520264 12520266 12520267 12520269 12528519 12528411 12530586 12530588 12530589 12530587 12531133 12531132 12532495 12532494 12532490 12532491 12532493 12532492 12533583 12533588 12533589 12533562 12533563 12533561 12533584 12533587 12533585 12533586 12533259 12533261 12533262 12533263 12534455 12534456 12534461 12540265 12540264 12544944 12546962 12547840 12547841 12549213 12552380 12552381 12553970 12553971 12553972 12553973 12553558 12553773 12553559 12553974 12553766 12553767 12553774 12555796 12555818 12555808 12555807 12563092 12563091 12563094 12564425 12568586 12568682 12572289 12572290 12572295 12572296 12572301 12572304 12572294 12572306 12572308 12572297 12572298 12572292 12572299 12572300 12572305 12572302 12572307 12572904 12572907 12572908 12572900 12572901 12572902 12572905 12572906 12572903 12575531 12579325 12582556 12582557 12583764 12583765 12583766 12583762 12583767 12583769 12583770 12583761 12583768 12583771 12586361…
    FDA recall number
    Z-0671-2025
    Quantity
    40,763

    The text above is the beginning of the record; 366,062 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • BD Pyxis CII Safe ES, REF: 1116-00

    Lot or code
    All software versions. REF/UDI-DI: 1116-00/10885403512605/ 15938609 15938608 15960426 15977244 15977243 15997587 15997584 15997586 16001276 16004167 16004170 16004168 16006959 16016028 16018286 16018285 16027068 16034363 16038606 16038608 16042040 16043100 16051041 16052300 16056892 16081629 16089006 16107354 16121844 16124176 16124177 16124175 16124174 16124178 16142019 16142717 16147368 16150604 16150317 16154380 16154381 16179910 16185906 16186402 16186403 16188916 16188917 16189876 16191858 16191855 16192478 16192476 16198735 16205805 16205708 16210899 16211121 16211303 16211474 16216310 16216322 16216171 16216317 16216324 16216170 16217014 16217009 16217010 16217006 16222804 16229923 16234971 16234972 16234970 16234973 16234969 16235009 16236311 16238237 16238264 16239968 16247326 16253283 16262739 16267647 16273818 16273822 16277384 16280705 16282121 16282137 16282082 16283296 16287656 16291326 16292340 16294019 16295348 16295346 16297236 16297224 16299107 16300479 16309949 16315507 16317206 16317204 16317205 16317207 16319700 16319670 16322458 16322446 16323719 16327551 16335942 16337405 16338265 16338264 16344600 16344608 16349184 16349205 16354606 16380050 16381506 16381558 16411952 16411951 16412194 16413752 16413747 16415153 16415200 16415590 16416557 16416531 16418522 16418188 16418187 16418507 16423221 16425773 16426437 16433986 16444194 16445004 16444237 16445017 16460539 16461440 16460710 16460551 16468172 16468090 16468208 16468173 16468174 16469932 16474654…
    FDA recall number
    Z-0672-2025
    Quantity
    218

    The text above is the beginning of the record; 522 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of OK, LA, FL, AR, VA, KY, PA, AL, NC, CA, MN, AK, TX, MI, ME, OH, OR, NY, IL, RI, GA, WI, NH, KS, CO, WA, MO, MD, AZ, SC, NV, WV, NM, IN, NJ, TN, DE, MT, MA, VT, DC, NE, CT, ID, WY, MS, UT, HI, IA, SD, ND, GU, MP, PR and the countries of CA, BR, JP, ES, NZ, AU, FR, AR, MX, IT, CL, TR, SA, MY, CH, IE, TH, SG, KW, BE, DE, PT, QA, MA, FI, AE, EG, BS, GB, BM, GR, BH, KR, MC, PL, LB, HK, IN, IL, MO, IQ, TN, NO, PH, CO, TW, CN, JO, OM.

Source

Issued by
U.S. Food and Drug Administration
Firm
CareFusion 303, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: CareFusion 303, Inc., Month: December 2024, Week 2024-W51. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95643 on December 18, 2024.

Other notices of CareFusion 303, Inc.