Baxter Healthcare Corporation recalls
47 notices naming Baxter Healthcare Corporation as the recalling firm, since November 20, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Baxter Healthcare Corporation: TruSystem 7000 U Operating Table and 1 more
Hazard (FDA): The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result…
FDAMedical devices
Baxter Healthcare Corporation: TruSystem 7500 surgical table and 4 more
Hazard (FDA): There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop…
FDAMedical devices
Baxter Healthcare Corporation: Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product…
Hazard (FDA): Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients…
FDAMedical devices
Contamination
Baxter Healthcare Corporation: Brand Name: VOLARA System Product Name: VOLARA
Hazard (FDA): Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
FDAMedical devices
Baxter Healthcare Corporation: PRO+ MRS SURFACE FOR VERSACARE and 2 more
Hazard (FDA): There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect…
FDAMedical devices
Baxter Healthcare Corporation: Dextrose Injection
Hazard (FDA): Presence of Particulate Matter: particulate matter identified as stainless steel particles in the solution.
FDAMedicines
Baxter Healthcare Corporation: 0.9% Sodium Chloride Injection USP 500 mL
Hazard (FDA): Presence of particulate matter: Particulates identified as fiberglass
FDAMedicines
Baxter Healthcare Corporation: Baxter Novum IQ Syringe Pump and 1 more
Hazard (FDA): A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
FDAMedical devices
Baxter Healthcare Corporation: Vancomycin Injection and 16 more
Hazard (FDA): CGMP Deviations
FDAMedicines
Baxter Healthcare Corporation: Cefazolin in Dextrose
Hazard (FDA): Presence of Particulate Matter
FDAMedicines
Baxter Healthcare Corporation: Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
Hazard (FDA): Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
FDAMedical devices
Baxter Healthcare Corporation: Hillrom VOLARA System P.CIRCUIT KIT and 3 more
Hazard (FDA): There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization…
FDAMedical devices
Baxter Healthcare Corporation: Adapter for Head Positioning
Hazard (FDA): The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating…
FDAMedical devices
Baxter Healthcare Corporation: Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number…
Hazard (FDA): Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system…
FDAMedical devices
Baxter Healthcare Corporation: Blue Ventilator Adapter Module
Hazard (FDA): Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or…
FDAMedical devices
Baxter Healthcare Corporation: Brand Name: Welch Allyn
Hazard (FDA): Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
FDAMedical devices
Baxter Healthcare Corporation: Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009…
Hazard (FDA): Pumps were released without full testing being performed, including occlusion alarm testing.
FDAMedical devices
Baxter Healthcare Corporation: Sterile Water for Irrigation
Hazard (FDA): Presence of particulate matter: plastic particles from the bottle rim were observed floating into the solution
FDAMedicines
Baxter Healthcare Corporation: SIGMA Spectrum Infusion Pump
Hazard (FDA): Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication…
FDAMedical devices
Baxter Healthcare Corporation: Welch Allyn INF BAG
Hazard (FDA): Replacement bladder kit for the Welch Allyn two-piece reusable blood pressure cuff may contain a Large Adult size bladder rather than a Thigh size bladder in the replacement kits.
FDAMedical devices
Baxter Healthcare Corporation: Oral Probe
Hazard (FDA): Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing…
FDAMedical devices
Baxter Healthcare Corporation: CONTINU-FLO Solution Set and 15 more
Hazard (FDA): IV sets may leak.
FDAMedical devices
Baxter Healthcare Corporation: Baster Continu-Flo Solution Set
Hazard (FDA): Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
FDAMedical devices
Baxter Healthcare Corporation: Novum IQ LVP and 2 more
Hazard (FDA): Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP…
FDAMedical devices
- FDABaxter Healthcare Corporation: Acetaminophen Injection 1000 mg/100 mL
- FDABaxter Healthcare Corporation: Spectrum IQ Infusion Pump
- FDABaxter Healthcare Corporation: Baxter Novum IQ Large Volume Pump
- FDABaxter Healthcare Corporation: Baxter SIGMA Spectrum Infusion Pump V6 and 1 more
- FDABaxter Healthcare Corporation: Hillrom Q-Link 13 component and 1 more
- FDABaxter Healthcare Corporation: 2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles
- FDABaxter Healthcare Corporation: Baxter Novum IQ Syringe Pump
- FDABaxter Healthcare Corporation: Affinity 4 Birthing Bed
- FDABaxter Healthcare Corporation: Baxter Novum IQ LVP INFUSION SYSTEM
- FDABaxter Healthcare Corporation: Welch Allyn Life2000 Ventilation system and 1 more
- FDABaxter Healthcare Corporation: Welch Allyn Connex Vital Signs Monitor and 2 more
- FDABaxter Healthcare Corporation: Welch Allyn Connex Integrated Wall System and 6 more
- FDABaxter Healthcare Corporation: Baxter NovumIQ Syringe INFUSION SYSTEM
- FDABaxter Healthcare Corporation: Progressa Bed Surfaces
- FDABaxter Healthcare Corporation: Baxter Operating table column TS7500 MOBIUS and 13 more
- FDABaxter Healthcare Corporation: Baxter Sigma Spectrum Infusion System and 1 more
- FDABaxter Healthcare Corporation: Baxter Solution Set with Duo-Vent Spike and 2 more
- FDABaxter Healthcare Corporation: Welch Allyn and 3 more
- FDABaxter Healthcare Corporation: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library and 1 more
- FDABaxter Healthcare Corporation: Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp and 10 more
- FDABaxter Healthcare Corporation: Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp and 9 more
- FDABaxter Healthcare Corporation: Baxter AK 98 Dialysis Machine
- FDABaxter Healthcare Corporation: Baxter Spectrum IQ Infusion System with Dose IQ Safety Software