What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
248 more codes are in the record below.
TruSystem 7500 surgical table, Model/Catalog Number: 4091000
- Lot or code
- Serial numbers: 108491706, 108726496, 108758657, 108858029, 108945825, 109083212, 109343910, 109684319, 110016886, 108557062, 108736192, 108773032, 108882327, 108976094, 109174333, 109370104, 109690845, 110016887, 108574826, 108736193, 108773033, 108882328, 108976095, 109174334, 109396256, 109690846, 110072583, 108578755, 108736194, 108842519, 108883564, 108976101, 109174335, 109464647, 109708813, 110084982, 108580717, 108736195, 108842523, 108883565, 109047445, 109174336, 109464648, 109708816, 110086138, 108580718, 108743656, 108842524, 108896687, 109047446, 109228237, 109464649, 109885369, 110144223, 108580720, 108743657, 108842525, 108896688, 109047447, 109228238, 109464650, 109896735, 110144224, 108604827, 108745163, 108842526, 108912913, 109047448, 109249973, 109488031, 109896736, 110144225, 108604828, 108745164, 108842679, 108912914, 109047451, 109249974, 109493243, 109896737, 110158878, 108604832, 108745165, 108842680, 108931731, 109047452, 109249989, 109500155, 109896738, 110158879, 108604833, 108745166, 108849690, 108931732, 109047460, 109301944, 109561673, 109933665, 110175899, 108726492, 108745180, 108849691, 108945772, 109047461, 109301945, 109561674, 109948691, 110175900, 108726493, 108745181, 108849694, 108945823, 109078531, 109301946, 109571317, 110007863, 110163294, 108726495, 108758656, 108849695, 108945824, 109078533, 109343482, 109571318, 110016883, 110163293, 110183861, 110220396, 110220395; UDI-DI number: 00887761968639
- FDA recall number
- Z-0350-2027
- Quantity
- 129 units
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
- Lot or code
- Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425
- FDA recall number
- Z-0351-2027
- Quantity
- 77 units
Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
- Lot or code
- Serial numbers: 108517025, 110115554, 110144247, 110188916, 109558596, 110139748, 110166201, 110188918, 109737582, 110139769, 110177678, 110192686, 110115551, 110144245, 110185616, 109889060; UDI-DI number: 00887761967526
- FDA recall number
- Z-0352-2027
- Quantity
- 16 units
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
- Lot or code
- Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519
- FDA recall number
- Z-0353-2027
- Quantity
- 62 units
TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087
- Lot or code
- Serial numbers: 107971925; UDI-DI number: 00887761974067
- FDA recall number
- Z-0354-2027
- Quantity
- 1 unit
Distribution (FDA): Worldwide - US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Baxter Healthcare Corporation (other notices of this firm)
- Title