What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
TruSystem 7000 U Operating Table (dV), Product Code 1723633
- Lot or code
- UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
- FDA recall number
- Z-0332-2027
- Quantity
- 62 units
Mobile column TruSystem 7500 U, Product Code 1730720
- Lot or code
- UDI-DI: 00887761974388; Serial number: 110183858
- FDA recall number
- Z-0333-2027
- Quantity
- 1 unit
Distribution (FDA): US Nationwide distribution in the states of IL, MO, TX.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Baxter Healthcare Corporation (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.