Why it was recalled
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Spectrum IQ Infusion Pump, Product Code 3570009
- Lot or code
- UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
- FDA recall number
- Z-2415-2025
- Quantity
- 6 units
Distribution (FDA): US distribution to states of: AZ, FL
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Baxter Healthcare Corporation (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.