What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Lot or code
- UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- FDA recall number
- Z-2548-2026
- Quantity
- 10540 units
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Lot or code
- UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- FDA recall number
- Z-2549-2026
- Quantity
- 0
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Lot or code
- UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- FDA recall number
- Z-2550-2026
- Quantity
- 16187 units
Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Lot or code
- UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- FDA recall number
- Z-2551-2026
- Quantity
- 13881 units
Distribution (FDA): Worldwide distribution - US Nationwide and the country of Canada.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Baxter Healthcare Corporation (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.