FDA recall 96311: GE Healthcare (China) Co., Ltd.: SIGNA Creator and 4 more

FDA recall 96311 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

13 more codes are in the record below.

  • SIGNA Creator, SIGNA Explorer, NMRI systems

    Lot or code
    1. Creator A. UDI-DI 00195278577221, System ID MR00160 MR00171 MR00189 THMR54503051 (Update 12/29/25 - New System ID MRR10626) (Update 1/28/26 - New System IDs MRR10269 850270138) (Update 2/19/26 - Addl System IDs MR00299¿ PHMRI24836517 12440397 13131449¿ PC0084) B. UDI-DI 00195278577238, System ID 083027026539324 5973MR0087 5973MR0091 5973MR0095 5973MR0096 5973MR0097 5973MR0100 5973MR0103 5973MR0104 5973MR0106 5973MR0107 5973MR0109 9830MR0112 9830MR0113 9830MR0114 9830MR0117 A199750201 IDMRI24467771 MR00082 MR00111 MR00120 MR00138 MR00140 MR00143 MR00152 MR00157 MR00174 MR00179 MR00180 MR00187 MR00191 MR00192 MR00195 MR00210 MR00229 MR00231 MR00236 MR00249 MR00251 MR00258 MR00263 MR00282 MRR1170 MRR1182 MRR2198 MRR2508 MRR2860 MRR2916 MRR2938 MRR3171 MRR3873A MRR3954 MRR4356 MRR4501 MRR4726 MRR493 MRR5138 MRR5332 MRR5494 MRR5684 MRR569A MRR6219 MRR673 MRR7117 MRR9626 MRYR2151 VNMRI24476142 VNMRI24513835 (Updated 12/29/25 - Additional System IDs MRR3679 , MR00322) (Updated 2/19/26 - Additional System IDs 83027280426725 83027280425025 MR00219 VNMRI24845558 THMRI25011164¿) C. UDI-DI 00840682113762, System ID 6529149 7361387 07143820 082427030212 082427040272 082427040285 082427060095 082427060119 082427070318 082427080107 082427080129 082427090384 082427090403 082427090419 082427090433 082427090434 082427090436 082427090448 082427100397 082427100418 082427110140 082427120326 082427120330 082427130173 082427160292 082427160307 082427160310 082427160318 082427160319 082427160343…
    FDA recall number
    Z-1239-2025
    Quantity
    2538

    The text above is the beginning of the record; 28,745 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Brivo MR355, NMRI system

    Lot or code
    Brivo MR355 (No UDI) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02)
    FDA recall number
    Z-1240-2025
    Quantity
    22
  • Optima MR360, NMRI system

    Lot or code
    A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01) (Update 1/28/26 - Additional System ID 82427190104)
    FDA recall number
    Z-1241-2025
    Quantity
    33
  • SIGNA MR355, SIGNA MR360, NMRI system

    Lot or code
    1. SIGNA MR355 A. UDI-DI 00840682144407 System IDs 082427010189 082427010193 082427010195 082427010200 082427010202 082427010205 082427010207 082427010208 082427010209 082427010210 082427020099 082427020120 082427020123 082427020127 082427030181 082427030194 082427030199 082427030205 082427030210 082427030213 082427030217 082427030221 082427030228 082427040273 082427040294 082427040303 082427040307 082427040316 082427040318 082427040323 082427060097 082427070236 082427070240 082427070244 082427070247 082427070255 082427070257 082427070279 082427070285 082427070299 082427070309 082427070312 082427070313 082427070316 082427070327 082427070330 082427070335 082427070339 082427070343 082427070344 082427070348 082427080113 082427080116 082427080118 082427080133 082427080135 082427080137 082427090359 082427090474 082427100436 082427100460 082427100461 082427100466 082427100481 082427100500 082427100506 082427110098 082427110106 082427110125 082427110126 082427110147 082427110153 082427120294 082427120297 082427120306 082427120315 082427120332 082427120336 082427120339 082427120348 082427120362 082427120367 082427120374 082427130130 082427140315 082427140332 082427160293 082427160303 082427160311 082427160312 082427160316 082427160327 082427160333 082427160341 082427160351 082427160352 082427160359 082427160364 082427160366 082427160367 082427160369 082427160374 082427160394 082427160397 082427160406 082427160416 082427160419 082427160421 082427170028 082427170032 082427180074…
    FDA recall number
    Z-1242-2025
    Quantity
    241

    The text above is the beginning of the record; 1,950 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • SIGNA MR380, NMRI system, Non-US Only

    Lot or code
    SIGNA MR380 UDI-DI 00195278361257 System IDs 082427020110 082427020116 082427040288 082427040293 082427040296 082427040309 082427070258 082427070264 082427070267 082427070268 082427070272 082427070275 082427070284 082427070286 082427070287 082427070288 082427070289 082427070292 082427070293 082427070294 082427070295 082427070303 082427070304 082427070305 082427070306 082427070307 082427070310 082427070311 082427070317 082427070320 082427070337 082427090417 082427090428 082427090431 082427090439 082427090462 082427090468 082427100424 082427100427 082427100428 082427100438 082427100443 082427100452 082427100453 082427100465 082427100467 082427100468 082427100470 082427100472 082427100488 082427100498 082427110124 082427110127 082427110131 082427110134 082427110136 082427110137 082427110139 082427110158 082427120299 082427120310 082427120312 082427120317 082427120318 082427120329 082427120347 082427130133 082427130135 082427130136 082427130137 082427130138 082427130142 082427130149 082427130154 082427130160 082427130161 082427130162 082427130163 082427130165 082427130166 082427130167 082427130168 082427130169 082427130171 082427130172 082427130178 082427130179 082427140219 082427140220 082427140221 082427140228 082427140230 082427140233 082427140236 082427140238 082427140239 082427140243 082427140244 082427140247 082427140253 082427140254 082427140255 082427140256 082427140260 082427140261 082427140262 082427140263 082427140264 082427140266 082427140269 082427140270 082427140271…
    FDA recall number
    Z-1243-2025
    Quantity
    188

    The text above is the beginning of the record; 1,069 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Domestic US distribution nationwide. International distribution worldwide.

Source

Issued by
U.S. Food and Drug Administration
Firm
GE Healthcare (China) Co., Ltd.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: GE Healthcare (China) Co., Ltd., Month: March 2025, Week 2025-W10. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96311 on March 5, 2025.

Other notices of GE Healthcare (China) Co., Ltd.