FDA recall 95704: GE Healthcare (China) Co., Ltd.: GE Healthcare Discovery RT and 16 more

FDA recall 95704 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

171 more codes are in the record below.

  • GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:

    Lot or code
    UDI/DI 00840682118699, Serial/Sales Order Numbers: CBDYG2400058HM, CBDYG2400057HM, CBDYG2400056HM, CBDYG2400051HM, CBDYG2400052HM, CBDYG2400040HM, CBDYG2400053HM, CBDYG2400060HM, CBDYG2400054HM, CBDYG2400055HM, CBDYG2400059HM,
    FDA recall number
    Z-0600-2025
    Quantity
    11 units
  • GE Healthcare Optima 520, Catalog number 5439126; X-ray/computed tomography systems

    Lot or code
    UDI/DI 00840682102568, Serial/Sales Order Numbers: CBDBG2400051HM, CBDBG2400048HM, CBDBG2400053HM, CBDBG2400049HM, CBDBG2400050HM, CBDBG2400052HM, CBDBG2400054HM, CBDBG2400055HM, CBDBG2400056HM, CBDBG2400057HM, CBDBG2400047HM
    FDA recall number
    Z-0601-2025
    Quantity
    11 units
  • GE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography system

    Lot or code
    UDI/DI 00840682146937, Serial/Sales Order Numbers: CBDXG2400054HM, CBDXG2400059HM, CBDXG2400055HM, CBDXG2400058HM, CBDXG2400057HM, CBDXG2400060HM, CBDXG2400061HM, CBDXG2400056HM
    FDA recall number
    Z-0602-2025
    Quantity
    8 units
  • GE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems

    Lot or code
    UDI/DI 00840682146159, Serial/Sales Order Numbers: CBDAG2400021HM, CBDAG2400023HM, CBDAG2400024HM, CBDAG2400020HM, CBDAG2400022HM
    FDA recall number
    Z-0603-2025
    Quantity
    5 units
  • GE Healthcare Revolution Advance, Model number 5849504; X-ray/computed tomography systems

    Lot or code
    UDI/DI 00195278385215, Serial/Sales Order Numbers: CBDLG2400031HM, CBDLG2400035HM, CBDLG2400034HM, CBDLG2400038HM, CBDLG2400036HM, CBDLG2400033HM, CBDLG2400037HM, CBDLG2400032HM, CBDLG2400039HM, CBDLG2400040HM
    FDA recall number
    Z-0604-2025
    Quantity
    10 units
  • GE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system

    Lot or code
    UDI/DI 00840682146173, Serial/Sales Order Numbers: CBDWG2400109HM, CBDWG2400103HM, CBDWG2400102HM, CBDWG2400096HM, CBDWG2400111HM, CBDWG2400106HM, CBDWG2400115HM, CBDWG2400113HM, CBDWG2400105HM, CBDWG2400112HM, CBDWG2400116HM, CBDWG2400114HM, CBDWG2400107HM, CBDWG2400104HM, CBDWG2400110HM, CBDWG2400108HM
    FDA recall number
    Z-0605-2025
    Quantity
    16 units
  • GE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278718129, Serial/Sales Order Numbers: CBFLG2400002HM, CBFLG2400005HM, CBFLG2400004HM, CBFLG2400003HM
    FDA recall number
    Z-0606-2025
    Quantity
    4 units
  • GE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278385215, Serial/Sales Order Numbers: CBDPG2400006HM, CBDPG2400005HM, CBDPG2400004HM, CBDPG2400007HM
    FDA recall number
    Z-0607-2025
    Quantity
    4 units
  • GE Healthcare Revolution EVO, Model Number 5454001-260; X-ray/computed tomography system

    Lot or code
    UDI/DI 00840682109796, Serial/Sales Order Numbers: CBDGG2400018HM, CBDGG2400017HM
    FDA recall number
    Z-0608-2025
    Quantity
    2 units
  • GE Healthcare Revolution Frontier, Model Numbers: 1) 5232085-213, 2) 5873009; X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278362230: 1) 5232085-213, Serial/Sales Order Numbers: CBDUG2400017HM, CBDUG2400016HM; 2) 5873009, Serial/Sales Order Numbers: CBDUG2400014HM
    FDA recall number
    Z-0609-2025
    Quantity
    3 units
  • GE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system

    Lot or code
    1) 6665000-2, UDI/DI 00840682146180, Serial/Sales Order Numbers: CBDMG2400130HM, CBDMG2400134HM; 2) 6670000-2, UDI/DI 00840682146180, Serial/Sales Order Numbers: CBDMG2400114HM, CBDMG2400127HM, CBDMG2400128HM, CBDMG2400129HM, CBDMG2400131HM, CBDMG2400132HM, CBDMG2400133HM, CBDMG2400135HM, CBDMG2400136HM, CBDMG2400137HM, CBDMG2400138HM, CBDMG2400139HM, CBDMG2400140HM, CBDMG2400141HM, CBDMG2400142HM, CBDMG2400143HM, CBDMG2400144HM, CBDMG2400145HM, CBDMG2400146HM, CBDMG2400147HM, CBDNG2400007HM, CBDNG2400010HM; UDI/DI 00195278686718, Serial/Sales Order Numbers: CBDOG2400032HM, CBDOG2400033HM, CBDOG2400034HM, CBDOG2400035HM; 3) 6670000-200, UDI/DI 00195278686718, Serial/Sales Order Numbers: CBFMG2400171HM, CBFMG2400188HM, CBFMG2400149HM, CBFMG2400160HM, CBFMG2400148HM, CBFMG2400173HM, CBFMG2400166HM, CBFMG2400185HM, CBFMG2400152HM, CBFMG2400156HM, CBFMG2400157HM, CBFMG2400158HM, CBFMG2400159HM, CBFMG2400163HM, CBFMG2400177HM, CBFMG2400178HM, CBFMG2400179HM, CBFMG2400180HM, CBFMG2400153HM, CBFMG2400154HM, CBFMG2400150HM, CBFMG2400182HM, CBFMG2400184HM, CBFMG2400161HM, CBFMG2400165HM, CBFMG2400172HM, CBFMG2400155HM, CBFMG2400162HM, CBFMG2400174HM, CBFMG2400176HM, CBFMG2400168HM, CBFMG2400151HM, CBFMG2400169HM, CBFMG2400170HM, CBFMG2400175HM, CBFMG2400181HM, CBFMG2400183HM, CBFMG2400186HM, CBFMG2400187HM, CBFMG2400164HM, CBFMG2400167HM; 4) 6670000-6, UDI/DI 00840682146180, Serial/Sales Order Numbers: MA10G2400033YC, MA10G2400032YC, MA10G2400034YC
    FDA recall number
    Z-0610-2025
    Quantity
    72 units
  • GE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system

    Lot or code
    UDI/DI 0019527856860, Serial/Sales Order Numbers: CBDOG2400029HM, CBDOG2400031HM, CBDOG2400030HM
    FDA recall number
    Z-0611-2025
    Quantity
    3 units
  • GE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278520852, Serial/Sales Order Numbers: CBFSG2400030HM, CBFSG2400032HM, CBFSG2400037HM, CBFSG2400034HM, CBFSG2400031HM, CBFSG2400035HM, CBFSG2400029HM, CBFSG2400028HM, CBFSG2400033HM, CBFSG2400036HM
    FDA recall number
    Z-0612-2025
    Quantity
    10 units
  • GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278466488, Serial/Sales Order Numbers: 5469298, 5471981, 5461542, 5469371 4616154
    FDA recall number
    Z-0613-2025
    Quantity
    5 units
  • GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite, BrightSpeed Elite Select; X-ray/computed tomography system

    Lot or code
    UDI/DI 00195278505798, Serial/Sales Order Numbers: 203364, 8102303, 5468897, 5469818
    FDA recall number
    Z-0614-2025
    Quantity
    4 units
  • GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA: China Only: Revolution Discovery CT

    Lot or code
    UDI/DI 00840682102490, Serial/Sales Order Numbers: 5476353, 5472107, 5472476
    FDA recall number
    Z-0615-2025
    Quantity
    3 units
  • GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80662DA: China Only: Discovery CT750HD

    Lot or code
    UDI/DI N/A, Serial/Sales Order Numbers: 5472442
    FDA recall number
    Z-0616-2025
    Quantity
    1 unit

Distribution (FDA): Worldwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
GE Healthcare (China) Co., Ltd.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: GE Healthcare (China) Co., Ltd., Month: December 2024, Week 2024-W50. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95704 on December 11, 2024.