What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
171 more codes are in the record below.
GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
- Lot or code
- UDI/DI 00840682118699, Serial/Sales Order Numbers: CBDYG2400058HM, CBDYG2400057HM, CBDYG2400056HM, CBDYG2400051HM, CBDYG2400052HM, CBDYG2400040HM, CBDYG2400053HM, CBDYG2400060HM, CBDYG2400054HM, CBDYG2400055HM, CBDYG2400059HM,
- FDA recall number
- Z-0600-2025
- Quantity
- 11 units
GE Healthcare Optima 520, Catalog number 5439126; X-ray/computed tomography systems
- Lot or code
- UDI/DI 00840682102568, Serial/Sales Order Numbers: CBDBG2400051HM, CBDBG2400048HM, CBDBG2400053HM, CBDBG2400049HM, CBDBG2400050HM, CBDBG2400052HM, CBDBG2400054HM, CBDBG2400055HM, CBDBG2400056HM, CBDBG2400057HM, CBDBG2400047HM
- FDA recall number
- Z-0601-2025
- Quantity
- 11 units
GE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography system
- Lot or code
- UDI/DI 00840682146937, Serial/Sales Order Numbers: CBDXG2400054HM, CBDXG2400059HM, CBDXG2400055HM, CBDXG2400058HM, CBDXG2400057HM, CBDXG2400060HM, CBDXG2400061HM, CBDXG2400056HM
- FDA recall number
- Z-0602-2025
- Quantity
- 8 units
GE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems
- Lot or code
- UDI/DI 00840682146159, Serial/Sales Order Numbers: CBDAG2400021HM, CBDAG2400023HM, CBDAG2400024HM, CBDAG2400020HM, CBDAG2400022HM
- FDA recall number
- Z-0603-2025
- Quantity
- 5 units
GE Healthcare Revolution Advance, Model number 5849504; X-ray/computed tomography systems
- Lot or code
- UDI/DI 00195278385215, Serial/Sales Order Numbers: CBDLG2400031HM, CBDLG2400035HM, CBDLG2400034HM, CBDLG2400038HM, CBDLG2400036HM, CBDLG2400033HM, CBDLG2400037HM, CBDLG2400032HM, CBDLG2400039HM, CBDLG2400040HM
- FDA recall number
- Z-0604-2025
- Quantity
- 10 units
GE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system
- Lot or code
- UDI/DI 00840682146173, Serial/Sales Order Numbers: CBDWG2400109HM, CBDWG2400103HM, CBDWG2400102HM, CBDWG2400096HM, CBDWG2400111HM, CBDWG2400106HM, CBDWG2400115HM, CBDWG2400113HM, CBDWG2400105HM, CBDWG2400112HM, CBDWG2400116HM, CBDWG2400114HM, CBDWG2400107HM, CBDWG2400104HM, CBDWG2400110HM, CBDWG2400108HM
- FDA recall number
- Z-0605-2025
- Quantity
- 16 units
GE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278718129, Serial/Sales Order Numbers: CBFLG2400002HM, CBFLG2400005HM, CBFLG2400004HM, CBFLG2400003HM
- FDA recall number
- Z-0606-2025
- Quantity
- 4 units
GE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278385215, Serial/Sales Order Numbers: CBDPG2400006HM, CBDPG2400005HM, CBDPG2400004HM, CBDPG2400007HM
- FDA recall number
- Z-0607-2025
- Quantity
- 4 units
GE Healthcare Revolution EVO, Model Number 5454001-260; X-ray/computed tomography system
- Lot or code
- UDI/DI 00840682109796, Serial/Sales Order Numbers: CBDGG2400018HM, CBDGG2400017HM
- FDA recall number
- Z-0608-2025
- Quantity
- 2 units
GE Healthcare Revolution Frontier, Model Numbers: 1) 5232085-213, 2) 5873009; X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278362230: 1) 5232085-213, Serial/Sales Order Numbers: CBDUG2400017HM, CBDUG2400016HM; 2) 5873009, Serial/Sales Order Numbers: CBDUG2400014HM
- FDA recall number
- Z-0609-2025
- Quantity
- 3 units
GE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system
- Lot or code
- 1) 6665000-2, UDI/DI 00840682146180, Serial/Sales Order Numbers: CBDMG2400130HM, CBDMG2400134HM; 2) 6670000-2, UDI/DI 00840682146180, Serial/Sales Order Numbers: CBDMG2400114HM, CBDMG2400127HM, CBDMG2400128HM, CBDMG2400129HM, CBDMG2400131HM, CBDMG2400132HM, CBDMG2400133HM, CBDMG2400135HM, CBDMG2400136HM, CBDMG2400137HM, CBDMG2400138HM, CBDMG2400139HM, CBDMG2400140HM, CBDMG2400141HM, CBDMG2400142HM, CBDMG2400143HM, CBDMG2400144HM, CBDMG2400145HM, CBDMG2400146HM, CBDMG2400147HM, CBDNG2400007HM, CBDNG2400010HM; UDI/DI 00195278686718, Serial/Sales Order Numbers: CBDOG2400032HM, CBDOG2400033HM, CBDOG2400034HM, CBDOG2400035HM; 3) 6670000-200, UDI/DI 00195278686718, Serial/Sales Order Numbers: CBFMG2400171HM, CBFMG2400188HM, CBFMG2400149HM, CBFMG2400160HM, CBFMG2400148HM, CBFMG2400173HM, CBFMG2400166HM, CBFMG2400185HM, CBFMG2400152HM, CBFMG2400156HM, CBFMG2400157HM, CBFMG2400158HM, CBFMG2400159HM, CBFMG2400163HM, CBFMG2400177HM, CBFMG2400178HM, CBFMG2400179HM, CBFMG2400180HM, CBFMG2400153HM, CBFMG2400154HM, CBFMG2400150HM, CBFMG2400182HM, CBFMG2400184HM, CBFMG2400161HM, CBFMG2400165HM, CBFMG2400172HM, CBFMG2400155HM, CBFMG2400162HM, CBFMG2400174HM, CBFMG2400176HM, CBFMG2400168HM, CBFMG2400151HM, CBFMG2400169HM, CBFMG2400170HM, CBFMG2400175HM, CBFMG2400181HM, CBFMG2400183HM, CBFMG2400186HM, CBFMG2400187HM, CBFMG2400164HM, CBFMG2400167HM; 4) 6670000-6, UDI/DI 00840682146180, Serial/Sales Order Numbers: MA10G2400033YC, MA10G2400032YC, MA10G2400034YC
- FDA recall number
- Z-0610-2025
- Quantity
- 72 units
GE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system
- Lot or code
- UDI/DI 0019527856860, Serial/Sales Order Numbers: CBDOG2400029HM, CBDOG2400031HM, CBDOG2400030HM
- FDA recall number
- Z-0611-2025
- Quantity
- 3 units
GE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278520852, Serial/Sales Order Numbers: CBFSG2400030HM, CBFSG2400032HM, CBFSG2400037HM, CBFSG2400034HM, CBFSG2400031HM, CBFSG2400035HM, CBFSG2400029HM, CBFSG2400028HM, CBFSG2400033HM, CBFSG2400036HM
- FDA recall number
- Z-0612-2025
- Quantity
- 10 units
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278466488, Serial/Sales Order Numbers: 5469298, 5471981, 5461542, 5469371 4616154
- FDA recall number
- Z-0613-2025
- Quantity
- 5 units
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite, BrightSpeed Elite Select; X-ray/computed tomography system
- Lot or code
- UDI/DI 00195278505798, Serial/Sales Order Numbers: 203364, 8102303, 5468897, 5469818
- FDA recall number
- Z-0614-2025
- Quantity
- 4 units
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA: China Only: Revolution Discovery CT
- Lot or code
- UDI/DI 00840682102490, Serial/Sales Order Numbers: 5476353, 5472107, 5472476
- FDA recall number
- Z-0615-2025
- Quantity
- 3 units
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80662DA: China Only: Discovery CT750HD
- Lot or code
- UDI/DI N/A, Serial/Sales Order Numbers: 5472442
- FDA recall number
- Z-0616-2025
- Quantity
- 1 unit
Distribution (FDA): Worldwide distribution.