Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 95674: GE Healthcare (China) Co., Ltd.: GE Proteus XR/A radiographic system

FDA recall 95674 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • GE Proteus XR/A radiographic system

    Lot or code
    UDI-DI: 00840682120777; System ID: 907443PROTEUS 251847RAD 256737PRO 251435PROT 251275RAD 251275RAD2 256486XR1 256429CR2 334732PXRA 256332RRM1 334418PRO 334418XRA 334670RAD 256487RAD 870898PROT 479BVXR 870541PROT6 870226PROT 870208RF 870208XR1 520498PR1 520498PR2 520901GERF1 928348PRO 928348PRO2 520MERPRO RADNETRAD352 310888PRO 818563DR 818461DR 707421NBSDI1 510797RM1 714537RM3 801774ER 909336RAD1 909336RAD2 760379RAD 213744PROTXR1 213744PROTXR2 310825PRO2 925370ORTHO1 925370ORTHO2 AVHRD22776 818360XR1 310374PRO 909788R2 831424RAD2 408885EDB 562493RM1 805418XR1 707429NBGV1 805652DR2 925210RAD1 303FCAUSHXR 970945CARB 720CRHIXR 970565RAD1 303320RAD1 303DHXR1 615812PROT 970945FL4 970BASRAD 970945FL4L 970624XR2 303788SEDXR 303FCLNWBXR 303730RM5 303HIPXR 303450ED 203466PROT 203319PROT 860224PROT 203688PROTLW 203852ED1 561965RAD1 813571ER2 813916XRA 352596PROM 352597RAD1 727FSEDXR1 954HCAERPRKM 850654PRO 305MCHDORAL 904264RAD1 850862PROT 333MHRX 727528RAD1 904230RM1 904399ER1 904399RADWEST 904928RAD 904MMRAD 904OPWESTER 813344PROTXR 321FCR1 305MCHWBRAD 904OPTCCRAD 850469PROT 352333PROTWD 291ERPRO 386774RM1 386774RM2 HCERPRO1 386446RM2 407359OVPRO OVERPRO 386PCHRAD3 561246RAD 727789PROT 850434BER 305MCPCPRO 813763BRPCER1 954476RAD4 954587RM1 386238FHR1 772340XR1 772SLMCER1 904819WGRAD 727341ER 727341PPER1 813282PRO1 813442XRA 863773PROT 404251P1 404686CRAD5 404686MOT2 706396COM 678207PROT1 912466PRO 912466PROT 403150 706257SFP 706335BR1 706335PRO1 229276CR3 678474PROT1 3109851…
    FDA recall number
    Z-0550-2025
    Quantity
    956 systems

    The text above is the beginning of the record; 8,947 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
GE Healthcare (China) Co., Ltd.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: GE Healthcare (China) Co., Ltd., Month: December 2024, Week 2024-W49. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95674 on December 4, 2024.

Other notices of GE Healthcare (China) Co., Ltd.