GE Healthcare (China) Co., Ltd. recalls
3 notices naming GE Healthcare (China) Co., Ltd. as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
GE Healthcare (China) Co., Ltd.: SIGNA Creator and 4 more
Hazard (FDA): For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to…
FDAMedical devices
GE Healthcare (China) Co., Ltd.: GE Healthcare Discovery RT and 16 more
Hazard (FDA): GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical…
FDAMedical devices
GE Healthcare (China) Co., Ltd.: GE Proteus XR/A radiographic system
Hazard (FDA): Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.