Siemens Healthcare Diagnostics, Inc. recalls
28 notices naming Siemens Healthcare Diagnostics, Inc. as the recalling firm, since November 20, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Siemens Healthcare Diagnostics, Inc.: Atellica CH Revised C-Reactive Protein
Hazard (FDA): The potential for falsely elevated results.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL and 1 more
Hazard (FDA): Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Urine Albumin REF: 11537225
Hazard (FDA): Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E)…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog…
Hazard (FDA): out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Syva Emit Disopyramide Assay Enzyme Immunoassay
Hazard (FDA): Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Enzymatic Hemoglobin A1c
Hazard (FDA): The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Siemens Dimension Tacrolimus Flex Reagent Cartridge
Hazard (FDA): A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
FDAMedical devices
Siemens Healthcare Diagnostics Inc: epoc BGEM BUN Test Card
Hazard (FDA): Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Hazard (FDA): Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Enzymatic Hemoglobin A1c
Hazard (FDA): A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Urine Albumin
Hazard (FDA): Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: IMMULITE 2000 Oak Mix Specific Allergen and 2 more
Hazard (FDA): The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Assay: IMMULITE 2000 Intact PTH
Hazard (FDA): The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Urinary/Cerebrospinal Fluid Protein
Hazard (FDA): Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.
FDAMedical devices
Siemens Healthcare Diagnostics Inc: epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Hazard (FDA): Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L…
FDAMedical devices
Siemens Healthcare Diagnostics Inc: RAPIDPoint 500 Systems Wash/Waste Cartridge
Hazard (FDA): Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: ADVIA¿ 120/2120/2120i SETpoint Calibrator and 3 more
Hazard (FDA): Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
FDAMedical devices
Contamination
Siemens Healthcare Diagnostics, Inc.: Product: Atellica CH Diluent - CONS 2 x 1.5L
Hazard (FDA): Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica IM Homocysteine Calibrator and 1 more
Hazard (FDA): Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
FDAMedical devices
Siemens Healthcare Diagnostics Inc: Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
Hazard (FDA): Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of…
FDAMedical devices
Siemens Healthcare Diagnostics Inc: epoc BGEM BUN Test Card and 1 more
Hazard (FDA): The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Atellica CH Revised C Reactive Protein
Hazard (FDA): Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: epoc NXS Host and 7 more
Hazard (FDA): Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to…
FDAMedical devices
Siemens Healthcare Diagnostics, Inc.: Magnesium Flex reagent cartridge
Hazard (FDA): Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in…
FDAMedical devices
- FDASiemens Healthcare Diagnostics, Inc.: Atellica¿ CH ¿2-Microglobulin
- FDASiemens Healthcare Diagnostics, Inc.: Atellica CI Analyzer. Catalog Numbers: 10947347.
- FDASiemens Healthcare Diagnostics, Inc.: Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate…
- FDASiemens Healthcare Diagnostics, Inc.: IMMULITE 2000 PSA assay