What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
- Lot or code
- Lot U1/UDI: 00842768001048
- FDA recall number
- Z-2975-2026
- Quantity
- 20 units
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
- Lot or code
- Lots: 2597, 2627, 2709/UDI: 00842768023293
- FDA recall number
- Z-2976-2026
- Quantity
- 55 units
Distribution (FDA): Worldwide - US Nationwide distribution in the state of CA and the country of Italy.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.