What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
3 more codes are in the record below.
Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
- Lot or code
- Siemens Material Number: 10995498. UDI-DI: 00630414598253 Kit lot numbers: 86236A74, 44206A75. ***Updated 5/8/26*** Additional Lots 25675A77, 44505A78, 52364A79.
- FDA recall number
- Z-1821-2025
- Quantity
- 2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
- Lot or code
- Siemens Material Number: 10310376. UDI-DI 00630414468198 Kit lot numbers: 86237A74, 44207A75. ***Updated 5/8/26*** Additional lots 25674A77 52363A79
- FDA recall number
- Z-1822-2025
- Quantity
- 1,072 packs (2,144 units) - 156 packs US, 916 packs OUS
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.