Why it was recalled

Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • epoc NXS Host; Siemens Material Number (SMN): 11413475 (US); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413475; UDI-DI: 00630414606095 Software Version Number: 4.14.9.
    FDA recall number
    Z-0734-2025
    Quantity
    6,738 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413497 (EU); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413497; UDI-DI: 00630414605760 Software Version Number: 4.14.9.
    FDA recall number
    Z-0735-2025
    Quantity
    3,011 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413498; UDI-DI: 00630414605814; Software Version Number: 4.14.9.
    FDA recall number
    Z-0736-2025
    Quantity
    670 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413506; UDI-DI: 00630414605821; Software Version Number: 4.14.9.
    FDA recall number
    Z-0737-2025
    Quantity
    320 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413517(MX) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413517; UDI-DI: 00630414605838; Software Version Number: 4.14.9.
    FDA recall number
    Z-0738-2025
    Quantity
    1 unit
  • epoc NXS Host; Siemens Material Number (SMN): 11413518 (ROW) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413518; UDI-DI: 00630414605678; Software Version Number: 4.14.9.
    FDA recall number
    Z-0739-2025
    Quantity
    2240 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413583(CN, KR) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413583; UDI-DI: 00630414612447; Software Version Number: 4.14.9.
    FDA recall number
    Z-0740-2025
    Quantity
    23 units
  • epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

    Lot or code
    Siemens Material Number: 11413879; UDI-DI: 00630414631028; Software Version Number: 4.14.9.
    FDA recall number
    Z-0741-2025
    Quantity
    276 units

Distribution (FDA): Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam

Source

Issued by
U.S. Food and Drug Administration
Firm
Siemens Healthcare Diagnostics, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Siemens Healthcare Diagnostics, Inc., Month: January 2025, Week 2025-W01. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95809 on January 1, 2025.

Other notices of Siemens Healthcare Diagnostics, Inc.